3.5 Cable Requirements

The following cables are necessary for standard installations, see also system drawings given in section
System Installation – with
8.2
W1 power supply cable for scanner (preinstalled with system connection box)
W3 trigger/alarm cable for linescanner (preinstalled with system connection box)
W8 from the junction box of the position indicator to the system connection box (standard installation) or
to TSM remote control box (when used with accessory Tire Slip Monitoring).
W9 from the TSM remote control box to the system connection box. This cable is used for data
communication, power supply, and trigger pulse.
W14 from the MI connection box to the system connection box and between MI connection boxes when
more than one shadow sensor is installed. This cable is used for data communication and power supply
to the pyrometer.
W15 from the Endurance connection box to the system connection box. This cable is used for data
communication and power supply to the pyrometer.
W16 MI sensing head cable.
W17 Endurance sensing head cable.
W18 Ethernet communication cable for scanner (preinstalled with system connection box)
W20 Fiber optic cable from system connection box to the control room and to the second scanner (if
used). Fiber optic cable to be provided by customer.
W22 Ethernet cable from fiber optic converter (control room) to the PC.
W23 from the position indicator head to the TSM junction box.
W24 Power supply cable 100 to 240 VAC.
W25 Fiber optic cable from the first scanner to the second scanner (if used). Fiber optic cable to be
provided by customer.
All copper cables must be shielded! The wires from W9, W14, and W15 must be a twisted pair! Local building
Cable
Distance
W1, W3
7.5 m (25 ft.)
W8
350 m (380 yd)
W9
350 m (380 yd)
W14
350 m (380 yd)
350 m (380 yd)
W15
350 m (380 yd)
W16
8 m (26 ft)
W17
15 m (49 ft.)
Accessories, page 48.
Note
codes should be observed when selecting cables!

Table 3-1: Required Cables

Cable features
preinstalled
3 x 0.25 mm²,
24 AWG, 3 conductors, shielded
2 x 2 x 0.25 mm²,
24 AWG, 3x twisted pair, shielded
2 x 2 x 0.5 mm²,
20 AWG, 2 x twisted pair, shielded
2 x 2 x 1.5 mm²,
16 AWG, 2 x twisted pair, shielded
2 x 2 x 1.5 mm²,
16 AWG, 2x twisted pair, shielded
preinstalled
preinstalled
Supplied from ...
Example / Remarks
Manufacturer
Customer
(N)YLHCY-J 3 × 0.25 mm²
Manhattan/CDT, P/N M13233
Customer
LifYCY 2 × 2 × 0.25 mm²
Customer
max. 5 pyrometers
Customer
max. 14 pyrometers
Customer
Manufacturer
Manufacturer
Pre-Installation
3
Cable Requirements
19