5 Unit Test Functions
5.1 General
After power-up and during operation, the Responder 1000/1100 runs automatic self-tests.
After power-up, the device beeps and all LED's will briefly light up.
When the unit detects a error, the "Service" LED lights up and an acoustic signal sounds.
Every three seconds the unit try's a new restart and test whether the error still appears.
5.2 Test Mode
The test mode is activated by inserting a jumper on PCB Responder 1000/1100 (J3 5-6). In the test
mode, some LED's of the energy levels (2J to 30J) will indicate any detected malfunction. The
table below lists all error codes and the corresponding explanations.
If the device detects a error, then:
1.)
switch Off the device
2.)
insert jumper J3 5-6 (use the jumper J2 3-4 for this test)
3.)
switch On the device
After power up, the "Service" LED lights up, an acoustic signal sounds and the error code will
indicated by the LED's of the energy levels. This mode is active until the device will switched
Off.
227 487 20 Rev G
Responder 1000/1100
Service Instruction
Page 18/38