8.3 Resistance Output

Maximum
Range
output range
400Ω
0~440Ω
3200Ω
400~3600Ω
-10 ℃~ 18 ℃, +28 ℃~ 55 ℃ temperature coefficient,±0.005%FS/ ℃ .

8.4 RTD

Graduation
Range
-200℃~
Pt100
850℃
-200℃~
Pt1000
650℃
Cu50
-50℃~150℃
Cu100
-50℃~150℃
As for exciting current during measurement, please refer to resistance measurement function.
As for allowable external exciting current during output, please refer to resistance output function.
2-wire: Does not include lead resistance.
3-wire: Assumes matched leads with a total resistance not exceeding 25Ω.
External exciting
Resolution
0.01Ω
0.4mA~4.0mA
0.1Ω
0.1mA~0.5mA
Resolution
2-wire
3-wire
0.7
0.4
0.1℃
1.2
0.7
current
ETX-1812
Accuracy (℃)
ETX-1812
4-wire
Output
0.4
0.7
0.3
0.3
0.8
0.8
0.4
0.4
Accuracy (Ω)
ETX-2012
0.15
0.25
0.50
1.0
ETX-2012
2-wire
4-wire
Output
3-wire
0.4
0.3
0.3
0.3
0.15
0.15
0.8
0.5
0.5
0.4
0.25
0.25