Advanced Application; Setting 0 % And 100 % Output Parameters - East Tester ET-2712 RTD User Manual

Table of Contents

Advertisement

6 Advanced Application

6.1 Setting 0 % and 100 % output
parameters
As for stepping operation and percentage display, 0% and 100% should be set before
using. Some level values have been set when it is out of the factory and set values are
shown as below:
Output function
Resistance 400Ω
Resistance 3200Ω
Pt1000
Cu100
The set values may not meet your requirements, so which can be reset. Then you can
display the percentage with stepping or slope output as well as display the percentage.
Adjust the output value with keys
works to enter into 0% or 100% values. The new defined range value is saved in the
calibrator storage automatically and remains effective after restart.
operations are available with the setting:
Press
on 25% increment.
Press
100%.
Pt100
Cu50
or
for manual steeping (increase or decrease) output based
or
instantly to make output jump within the range of 0% to
0% value
0.00 Ω
0.0 Ω
0.0 ℃
0.0 ℃
0.0 ℃
0.0 ℃
, press
or
100% value
400.00 Ω
3200.0 Ω
500.0 ℃
400.0 ℃
150.0 ℃
150.0 ℃
for a long time until the buzzer
The flowing
14