4 Advanced Application

4.1 Setting 0 % and 100 % output
parameters
As for stepping operation and percentage display, 0% and 100% should be set before
using. The default values have been set when delivered from the factory and set values
are listed below:
Frequency – 200Hz
Frequency – 2000Hz
Frequency – 20kHz
The default set values may not meet your requirements, so you can reset them. Press
and hold
or
reset value will be stored in the storage space of the calibrator automatically and
remains effective after restart. Now you can start operation with the reset value:
Manually stepping an output with 25 % increments.
Jump between the 0 and 100 % span points by momentarily pushing
.
Source function
DC V
DC mV
DC mA
to reset 0% and 100% value until the buzzer works. The
0%
0.000 V
10.000 V
0.00 mV
100.00 mV
4 .000 mA
20.000 mA
0.00 Hz
200.00 Hz
0.0 Hz
2000.0 Hz
0 .000 kHz
20 .000 kHz
100%
or
15