NOTE:
• The pipettor is designed for the transfer of liquids only
using the tip. Do not aspirate liquids without the tip
attached. The aspirated liquid should not enter the
pipettor, as it may cause damage.
• Single-use tips reduce the risk of contamination
of samples.
• Keep the pipettor clean, avoiding the use of abrasive
or corrosive cleaning agents (e.g., acetone).
• Keep the pipettor upright when there is liquid in
the tip.
• Only using the pipettor in accordance with the manu-
facturer's instructions ensures the correct pipettor
parameters are maintained.
• After replacing the plunger or the shaft, the pipettor
should be calibrated.
• In the case of incorrect operation, the device should
be cleaned in accordance with the Instructions for
Use or transferred to a service point.
• Ambient operating temperature is +5°C to 45°C.
• Ambient storage conditions (in the original packaging
during transport and short storage) is -25°C to 55°C.
When working with the pipettor:
• Follow general work safety regulations regarding
hazards related to work in the laboratory.
• Take special care when pipetting aggressive
substances.
• Use appropriate protective attire (e.g., clothing,
goggles, and gloves).
• Avoid pointing the pipettor at yourself or others
during use.
• Only use parts and accessories recommended
by the manufacturer.

5. Specifications

The pipettor is a high quality instrument which offers
excellent accuracy and precision. The accuracy and
precision (repeatability) of the liquid volume depends on
the quality of pipet tips used. The values for accuracy,
and precision shown in the table below were obtained
using manufacturer non-filter pipet tips. Those tips are
recommended for use to ensure compatibility, accuracy,
and precision when pipetting.
6 – VWR UHP PIPETTORS
The accuracy and precision were obtained gravi-
metrically using manufacturer tips performing at least
10 measurements of distilled water at a temperature of
20°C ± 1°C according to EN ISO 8655 standards.
The use of tips from other manufacturers or filter tips
may result in incorrect liquid aspiration and require
pipettor recalibration.
Nominal
Volume
Volume
Accuracy
(μL)
(μL)
(%)
Single-channel Pipettors
Min.
0.1
±40.0
0.2
±12.0
2
1
±2.7
Max.
2
±1.5
Min.
0.5
±4.0
1
±2.5
10
5
±1.5
Max.
10
±0.5
Min.
2
±3.0
20
10
±1.0
Max.
20
±0.8
Min.
5
±2.5
50
25
±1.0
Max.
50
±0.8
Min.
10
±1.6
100
50
±1.0
Max. 100
±0.8
Min.
20
±1.2
200
100
±1.0
Max. 200
±0.6
Min. 100
±1.6
1,000
500
±0.8
Max. 1,000
±0.6
Permissible
Preci-
Non-
Volumes
sion
filter
(μL)
(%)
Tips (μL)
0.06-0.14
≤12.0
0.176-0.224
≤6.0
10
0.973-1.027
≤1.3
1.97-2.03
≤0.7
0.48-0.52
≤2.8
0.975-1.025
≤1.8
10
4.925-5.075
≤0.7
9.9-10.1
≤0.4
1.93-2.07
≤1.5
9.9-10.1
≤0.5
200
19.84-20.16
≤0.3
4.875-5.125
≤2.0
24.75-25.25
≤0.6
200
49.6-50.4
≤0.4
9.7-10.3
≤0.8
99.5-50.5
≤0.3
200
99.2-100.8
≤0.2
19.6-20.4
≤0.6
99.0-101.0
≤0.3
200
198.8-201.2
≤0.2
97.5-102.5
≤0.4
496-504
≤0.3
1,000
994-1,006
≤0.15
Instruction Manual – 7