5.5

X-Ray Generation Limit

Tube Voltage (kV)
60 ~ 65
61 ~ 70
66 ~ 70
60
66
67
68
69
70
EzRay Premium User Manual
Tube Voltage, Tube Current, and Exposure Time value cannot be set
to the maximum value simultaneously.
If the parameter you set exceeds the safety limit, the "Over Value"
message is displayed on the top left of the LCD. Unless this message
is cleared, you cannot perform irradiation.
Tube Current (mA)
Adjustable step: 1 kV, 1 mA, 0.01 sec
5. Using the System
Exposure Time (sec)
4 ~ 6
7
4 ~ 5
7
6
6
6
6
6
0.04 ~ 2.0
Not available
0.04 ~ 2.0
0.04 ~ 0.5
0.04 ~ 1.8
0.04 ~ 1.5
0.04 ~ 1.2
0.04 ~ 0.9
0.04 ~ 0.5