CLEANING AND DISINFECTION
To avoid pinch risk of hand / fi ngers, IndiGo must be deactivated and brakes on the
bed applied prior to removing obstructions, cleaning and / or servicing the IndiGo wheel.
For cleaning and disinfection, please refer to cleaning and disinfection instructions in your bed IFU.
CARE AND PREVENTIVE MAINTENANCE
Actions To Be Completed By Caregiver
Check that the IndiGo cover is undamaged
Check that IndiGo is activated when pedal is in its most upward position
Check the emergency stop assemblies for signs of damage
Check that applicable IndiGo labeling is present and legible
If the result of any of these tests is unsatisfactory, do not use IndiGo. Contact Arjo
or an Arjo-approved service agent.
The procedure below must be carried out by suitably trained and qualifi ed personnel.
Failure to do so may result in injury or an unsafe product.
General
Examine the IndiGo for obvious signs of damage.
Exercise the unit's operation, including the emergency stop switches. If any abnormalities
are observed, refer to the Troubleshooting chapter.
Examine the IndiGo cables for cuts, abrasions, kinks or other deterioration. Contact Arjo
if damage is observed. Refer to Contact information in the back of this manual.
Examine the IndiGo drive wheel and tire assembly for damage. If abnormalities are observed,
replace the Drive Wheel Assembly. Refer to Drive Wheel Replacement procedures in the Servicing
chapter to replace.
Visually inspect the gas spring for excessive oil on shaft. Ensure the gas spring maintains downward
force on the drive wheel when IndiGo is engaged. If excessive oil is present on the shaft or the gas
spring does not maintain downward force when IndiGo is engaged, replace the gas spring. Refer to
Gas Spring Replacement procedure in the Service Manual.
Check all accessible nuts, bolts and other fasteners to ensure they are present and correctly
tightened.
Check that IndiGo does not interfere with the scale function through the bed's full range of motion.
Daily
Weekly
13
Need help?
Do you have a question about the Assembly instructions and is the answer not in the manual?