ELECTRICAL MAGNETIC COMPATIBILITY INFORMATION
The use of non-approved accessories may result in increased emissions or reduced
immunity of the equipment. A list of approved accessories is included in the product
Instructions for Use.
When this equipment is used adjacent to other electronic devices, the user should
observe the equipment to verify normal operation.
Guidance and manufacturer's declaration – electromagnetic emissions
IndiGo is intended for use in the electromagnetic environment speci ed below. The customer or the user
of the IndiGo should ensure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage uctuations /
icker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
IndiGo is intended for use in the electromagnetic environment speci ed below. The customer or the user
of IndiGo should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Compliance
Electromagnetic environment - guidance
IndiGo uses RF energy only for its internal function.
Group 1
Therefore its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
Class A
IndiGo is suitable for use in all establishments other than
domestic and those directly connected to the public low-
Class A
voltage power supply network that supplies buildings used
for domestic purposes.
Complies
IEC 60601
Compliance
test level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for power
supply lines
± 2 kV
± 1 kV for input/
Not applicable
output lines
± 1 kV line to line
± 1 kV
± 2 kV line to
± 2 kV
earth
Electromagnetic environment - guidance
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
19