Contents Safety instructions Page For your protection • Read the operating instructions in full before starting up and follow the Declaration of conformity safety instructions. Safety instructions • Keep the operating instructions in a place where they can be accessed by Unpacking everyone.
Page 3
The safe temperature limit must always be set to at least 25 °C lower than the • It may be possible for wear debris from rotating accessory parts to reach the fire point of the media used. material being processed. •...
Unpacking Assembling the stand • Unpacking Remove screw plugs (M) - Please unpack the device carefully Remove the protective cap from the support rod - In the case of any damage a detailed report must be set immediately Put the washer between housing and nut (post, rail or forwarder) Screw the support rod onto the device by hand until the end stop is reached...
Operation Ensure that the protective film is removed from the base plate before use! Move device switch (A) to the OFF position Commissioning Insert the mains power cable into the power socket (K) Move device switch (A) to the ON position Operating mode A is preset (see “Operating modes”) Adjust the motor speed using the rotary knob (D) The value selected will be shown on the display (F)
Setting operating mode Er 5 Operating the device in mode A, B or D Providing safety protection, Error 5 is taken to recognise that the sensor has not been placed in the medium when the heating is switched on. Mode A All settings will be stored if the device is switched off or disconnected from the power supply.
• Turn the “Safe Temperature” setting screw slowly anticlockwise until the Setting the safe temperature limit heating function switches off and the display shows Er 25 . • The “Safe Temp” value is displayed on the screen. The maximum achievable heating plate temperature is restricted by an adjustable safety temperature limit.
Please send instrument in for repair only after it has been cleaned and is free from any materials which may constitute a health hazard. For this you should request the “Decontamination Clearance Certificate” from IKA , or use the download printout of it from the IKA website ® ®...
Error codes Any malfunctions during operation will be identified by an error Switch off device using the main switch (A) message on the display (E and F). Carry out corrective measures Proceed as follows in such cases: Restart device Error code Cause Effect Solution Temperature inside device too high...
Technical Data Number of stirring positions Connection for ext. temperature sensor PT 1000 Stirring quantity max. per stirring position (H Temperature setting resolution of medium Motor rating output Adjustable safety circuit °C 50 - 360 Direction of rotation right Set-up plate material stainless steel 1.4301 Speed display set-value...
USB interface Analog output Voltage 220 - 230 Frequency 50/60 Power input Subject to technical changes! Warranty In accordance with ® warranty conditions, the warranty period is 24 months. The warranty does not cover worn out parts, nor does it apply to faults resulting For claims under the warranty please contact your local dealer.
Need help?
Do you have a question about the RET basic and is the answer not in the manual?