Professional Series
2.13
Reset to Factory Default Settings
Resetting the unit to Factory Default Settings will do the following:
•
Clear all single point calibration (SPC) temperatures.
•
Return TLIM to the unit's maximum allowable temperature.
•
Turn off Power Recovery and the temperature dependent timer.
•
Re-enable the beeper setting.
1. To reset the unit to factory default settings, the unit
must first be in the top level of the settings menu.
2. Rotate the left knob (C) to scroll to the 'SYS'
(System) setting.
3. Briefly press the left knob (C) to enter the System
settings menu.
4. Rotate the left knob (C) to the "RSET" (Reset)
setting.
5. Briefly press the left knob (C) to change the Reset
settings.
The current Reset setting will begin to blink.
6. Rotate the left knob (C) to scroll to the desired
Reset setting.
7. Press and hold the left knob (C) until the unit
beeps to confirm the Reset setting.
8. Rotate the left knob (C) to the "BACK" icon.