Using The Temperature Limitation Feature - VWR Professional Series Instruction Manual

Hotplate-stirrers
Hide thumbs Also See for Professional Series:
Table of Contents

Advertisement

Professional Series

2.8 Using the Temperature Limitation Feature

Temperature Limitation (TLIM) allows the user to
change the maximum temperature limit of the heater.
1. To control the TLIM Feature, the unit must first be
in the top level of the Settings menu.
2. Rotate the left knob (C) to scroll to the "TLIM"
feature.
3. Briefly press the left knob (C) to change the
temperature limit of the heater.
The temperature will begin to blink to indicate that
it can be modified.
4. Rotate the left knob (C) to the desired temperature
limit of the heater.
5. Briefly press the left knob (C) to confirm the
setting.
The temperature will stop blinking.
6. To exit the settings menu, rotate the left knob (C)
to scroll to 'EXIT'.
7. Briefly press the left knob (C) to return to the main
operating screen.
8. The TLIM icon (V) will be illuminated to indicate
that there is a temperature limit that is less than
the default setting.
The heat setting (M) will not be able to be set
above the set temperature limit.
Note: While using the external probe, TLIM will still
limit the temperature of the heater. This may cause
the sample to heat at a slower rate and peak at a
lower temperature than the TLIM setting.