3. Installation and commissioning

The device was thoroughly cleaned before leaving the factory. However, due to the transport con-
ditions and the transport packaging, neither a sterile nor a disinfected product can be assumed
EN
upon delivery. For the initial cleaning, please observe the instructions provided in
Section
5.1.
First check the device and accessories for completeness by comparing with the enclosed delivery
bill.
The device may only be operated in an upright position; operation with an inclination of more than
30° from the normal in one direction is not permitted. Operation in inadmissible positions (upside
down on its head, lying on its back or on its display) is not permitted and may result in damage to
the device.
Only use the screws supplied. Always use all fastening points for safe mounting of the device.
Adequate ventilation is necessary for reliable and continuous operation of the device. Ventilation
protects the device from overheating. Therefore, do not install the device in a location or under
conditions where adequate ventilation of the housing surfaces is not possible. To avoid heat accu-
mulation or interruption of air circulation and to protect the device mounting from possible over-
loads, it is not permitted to place objects or materials on the device or to hang them on the device.
To avoid the risk of injury and damage to the device, all connection cables should be routed so that
no one can step on them, trip over them, or place heavy objects on them. When mounting, ensure
a fixed position and/or stable suspension.

3.1 Cable and device connection

Use only the supplied connection cables.
First connect the potential equalization cable (special green-yellow marking) to ensure sufficient
potential equalization between user and device. Operation in the medical field without connected
potential equalization is not permitted.
Before connecting to a supply network, check the correct voltage and frequency of your power
supply network (see nameplate on the housing). Carefully connect all desired connection cables
(power cable, keyboard, mouse, serial cable, network, etc.) to the device. To avoid the risk of elec-
tric shock, this device must only be connected to a supply network with a protective earth conduc-
tor. Finally, check the tight fit of all cable connections. It is not necessary for the end customer to
attach strain reliefs to the device.
For complete disconnection from the mains, the system mains switch must be actuated or the
mains plug must be pulled out of the socket. When disconnecting from the mains, always grasp
the mains cable by the plug. Do not pull on the cable. Do not place the device in conditions that
make it difficult to disconnect it from the mains power supply.