Dc Current Source Verification; Dc Current Measure Verification - Fluke 71 Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71 Series:
Table of Contents

Advertisement

Table 56. 715 DC Voltage Source Verification (0.000 to 20.000 V)
Fluke 715
0.000 V
5.000 V
15.000 V
5. Press the V key on the 715 Calibrator. Display should change to:
OUTPUT
6. Using the large [ key scroll button of the 715 Calibrator, press to step to the
voltages in Table 57, verifying the output on the Fluke 8508A.
Table 57. 715 DC Voltage Source Verification (0.00 to 200.00 mV)
Fluke 715
0.00 mV
50.00 mV
150.00 mV

DC Current Source Verification

1. Disconnect the Fluke 8508A. Press the A key on the 715 Calibrator. The display
should change to:
OUTPUT
2. Connect test leads from the 715 Calibrator's mA output jacks (black to V jack and red
to +LOOP jack) to the Fluke 8508A input mA jacks (black to I- and red to I+).
3. Set the Fluke 8508A function to [
the 715 Calibrator, press to step to the current outputs in
Table 58, verifying the readings on the Fluke 8508A.
Fluke 715
4.000 mA
12.000 mA
24.000 mA

DC Current Measure Verification

1. Press I key on the 715 Calibrator. Display should indicate closely to:
INPUT
2. Press the A key and the display should change to:
INPUT
3. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
terminal.
4. Connect a test lead from Black AUX terminal of the 5520A to the 715 Calibrator
Com terminal.
5. Connect a test lead from I- terminal of the Fluke 8508A to the 715 Calibrator mA
terminal.
0.00 mV
0.000 mA
DC CURRENT
Table 58. 715 DC Current Source Verification
-25.00 mA %
0.000 mA
Fluke 8508A
-0.002 V to + 0.002 V
4.998 V to 5.002 V
14.997 V to 15.003 V
Fluke 8508A
-0.02 mV to + 0.02 mV
49.98 mV to 50.02 mV
149.96 mV to 150.04 mV
]. Using the large [ scroll button of
Fluke 8508A
3.9976 mA to 4.0024 mA
11.9968 mA to 12.0032 mA
23.9956 mA to 24.0044 mA
Process Calibrators
Verification
37