Ma/Volts Source Measure - Fluke 71 Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71 Series:
Table of Contents

Advertisement

71X Series
Calibration Manual
5. Press any key on the 715 Calibrator. The display should read:
6. Set the Fluke 5520A to 24.0000 mA.
7. Press any key on the 715 Calibrator. The display should read:
8. Connect test lead from the NORMAL output jacks of the Fluke 5520A to the voltage
9. Set the Fluke 5520A to 0.0000 mV and press O.
10. Press any key on the 715. A relay will click and the display should read:
11. Set the Fluke 5520A to 10.00000 V.
12. Press any key on the 715 Calibrator. The display should read:
13. Set the Fluke 5520A to 0.0000 mV.
14. Press any key on the 715 Calibrator. The display should read:
15. Set the Fluke 5520A to 100.0000 mV.
16. Press any key on the 715 Calibrator. The display should read:

mA/Volts Source Measure

1. Remove all test leads from the 715 Calibrator.
2. Press any key on the 715 Calibrator. Display should change to:
3. Take one single test lead and plug one end into the V jack and the other end into the
56
INPUT
-----
mA
then change to:
INPUT
24.000 mA
INPUT
-----
mA
then change to:
INPUT
0.000 V
jacks on the 715 Calibrator (black to COM jack and red to the V jack).
INPUT
-----
V
then change to:
INPUT 10.000 V
INPUT
-----
V
then change to:
INPUT
0.00 mV
INPUT
-----
mV
then change to:
INPUT
100.00 mV
INPUT
-----
mV
then change to:
OUTPUT
mV
OUTPUT ----- mV
After a couple of seconds, display should change to:
----- V
OUTPUT
After a couple more seconds, display should change to:
OUTPUT
mA
+LOOP jack of the 715 Calibrator (places a short between the two jacks).