Specifications - Thermo Scientific 8600 Series Operating And Maintenance Manual

86c ultra low temperature freezer
Table of Contents

Advertisement

Model
8602
Temperature
-50°C (-58°F) to -86°C (-123°F) in an 18C to 32C* (64.4F to 89.6F) ambient
Range
33.3"W x 77.8"H x 31.0"
Exterior
Dimensions
84.6 x 197.6 x 78.7 cm
23.0"W x 51.5"H x 19.3"
Interior
Dimensions
58.4 x 130.8 x 49.0 cm
13.0 cu. ft.
Capacity
(368.1 liters)
Refrigeration Two 1 HP (2545 BTUH each)
Insulation
Non-CFC, foamed-in-place urethane: 5.0" (12.7 cm) cabinet; 4.5" (11.4 cm) door
230V, 50/60Hz, 12.0 FLA
Electrical
Operating Range:
208VAC-240VAC
15 amp, 230V,
Breaker
Dedicated Circuit,
Requirements
15 Amp
Time Delay Breaker
Shipping
712 lbs. (323.0 kg)
Weight Motor
* Compressors may not cycle off with cabinet running at -86C in a 32C ambient.
Thermo Scientific

Specifications

Section 7
8603
8604
33.3"W x 77.8"H x 31.0"
33.3"W x 77.8"H x 37.0"
84.6 x 197.6 x 78.7 cm
84.6 x 197.6 x 94.0 cm
23.0"W x 51.5"H x 19.3"
23.0"W x 51.5"H x 25.3"
58.4 x 130.8 x 49.0 cm
58.4 x 130.8 x 64.3 cm
13.0 cu. ft.
17.3 cu. ft.
(368.1 liters)
(489.9 liters)
120V, 60 Hz, 16.0 FLA
120V, 60 Hz, 16.0 FLA
Operating Range:
Operating Range:
108-130V
108-130V
20 amp, 120V,
20 amp, 120V,
Dedicated Circuit,
Dedicated Circuit,
20 amp
20 amp
Time Delay Breaker
Time Delay Breaker
712 lbs. (323.0 kg)
795 lbs. (360.6 kg)
8605
8606
33.3"W x 77.8"H x 37.0"
40.8"W x 77.8"H x 37.0"
84.6 x 197.6 x 94.0 cm
103.6 x 197.6 x 94.0 cm
23.0"W x 51.5"H x 25.3"
30.6"W x 51.5"H x 25.3"
58.4 x 130.8 x 64.3 cm
77.7 x 130.8 x 64.3 cm
17.3 cu. ft.
23.0 cu. ft.
(489.9 liters)
(651.3 liters
230V, 50/60Hz, 12.0 FLA
230V, 50/60Hz, 12.0 FLA
Operating Range:
Operating Range:
208VAC-240VAC
208VAC-240VAC
15 amp, 230V,
15 amp, 230V,
Dedicated Circuit,
Dedicated Circuit,
15 Amp
15 Amp
Time Delay Breaker
Time Delay Breaker
795 lbs. (360.6 kg)
900 lbs. (408.2 kg)
Model 8600 Series
7-1

Hide quick links:

Advertisement

Table of Contents
loading