Single Door Units
Model (Catalog No.)
Temperature Range
Exterior Dimensions
Interior Dimensions
Capacity
Refrigeration
Insulation
Electrical - nominal
voltage ±10%
Breaker Requirements
Shipping Weight
Model (Catalog No.)
Temperature Range
Exterior Dimensions
Interior Dimensions
Capacity
Refrigeration
Insulation
Electrical- nominal
voltage ±10%
Breaker Requirements
Shipping Weight
Thermo Scientific
Section 6
902 (902)/5902 (902SS)
-50°C (-58°F) to -86°C (-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient
33.3"W x 77.9" H x 32.9" F-B
(84.6cm x 197.9cm x 84.6cm)
23.0"W x 51.5"H x 19.3" F-B
(58.4cm x 130.8cm x 49.0cm)
13.0 cu. ft. (368.1 liters)
Two 1 HP (2545 BTUH each)
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 4.5" (11.4 cm) door
230VAC, 50 Hz, 12.0 FLA
Operating Range: 208VAC-240VAC
15 Amp, Dedicated Circuit,
15 Amp Time Delay Breaker
696 lbs. ( 316 kg)
904 (904)/ 5904 (904SS)
-50°C(-58°F) to -86°C(-123°F) in an 18C to 32C * (64.4F to 89.6F) ambient
33.3"W x 77.9" H x 38.9" F-B
(84.6cm x 197.9cm x 94.0cm)
23.0"W x 51.5"H x 25.3" F-B
(58.4cm x 130.8cm x 64.3cm)
17.3 cu. ft. (489.9 liters)
Two 1 HP (2545 BTUH each)
Non CFC, foamed-in-place urethane: 5.0" (12.7cm) cabinet; 4.5" (11.4 cm) door
120VAC, 60 Hz, 16.0 FLA
Operating Range: 108VAC-130VAC
20 Amp, Dedicated Circuit,
20 Amp Time Delay Breaker
763 lbs. ( 346 kg)
Specifications
903 (903)/5903 (903SS)
120VAC, 60 Hz, 16.0 FLA
Operating Range: 108VAC-130VAC
20 Amp, Dedicated Circuit,
20 Amp Time Delay Breaker
905 (905)/5905 (905SS)
230VAC, 50 Hz, 12.0 FLA
Operating Range: 208VAC-240VAC
15 Amp, Dedicated Circuit,
15 Amp Time Delay Breaker
909 (909)/5909 (909SS)
208-230VAC, 60 Hz, 12.0 FLA
Operating Range: 208VAC-240VAC
15 Amp, Dedicated Circuit,
15 Amp Time Delay Breaker
910 (910)/5910 (910SS)
208-230VAC, 60 Hz, 12.0 FLA
Operating Range: 208VAC-240VAC
15 Amp, Dedicated Circuit,
15 Amp Time Delay Breaker
Model 900 and 5900 Series