Section 2
Calibrate
Calibrate the Optional
Sample Probe
2-2
Model 8600 Series
For freezers with the optional sample probe, place the calibrated
instrument in the center of the sample bottle. The bottle should contain an
appropriate medium and the measuring instrument should be centered in
the bottle.
1. Press the Mode key until the Calibrate indicator lights.
2. Press the right arrow until "SAMP T = -XX.X" appears in the message
center.
3. Press up/down arrow to match display to calibrated instrument.
4. Press Enter to store calibration.
5. Press the Mode key to return to Run or the right/left arrow to go to
next/previous parameter.
See Chart 2-1 for calibration process functions.
Temperature Stabilization Periods
Startup - Allow 12 hours for the temperature in the cabinet to stabilize
before proceeding.
Already Operating - Allow at least 2 hours after the display reaches set
point for temperature to stabilize before proceeding.
Note During calibration, the temperature display is not available.
If no keys are pressed for approximately five minutes while in Calibration
mode, the system will reset to Run mode.