Replace The Battery(S) - Thermo Scientific 8600 Series Operating And Maintenance Manual

86c ultra low temperature freezer
Table of Contents

Advertisement

Vacuum Relief Port
(continued)
Replace the
Battery(s)
Thermo Scientific
Observe the inner side of port periodically for frost and ice build-up.
Remove any frost with a soft dry cloth. If the tube should become clogged
with ice, it must be cleaned. Clear the vacuum relief tube completely free
of ice to prevent rapid ice formation.
Factors that can affect the performance of the vacuum relief port include
high ambient temperature, high humidity conditions and frequent door
openings. Perform maintenance weekly, or as needed.
Note Failure to maintain the vacuum relief port may result in excessive ice
build-up inside the tube, clogging the port, and inability to open the door.
The vacuum relief port may need to be cleaned more often with frequent
door openings and high humidity environments.
1. To gain access to the battery, open the lower door by grasping the
bottom left corner. The battery is rectangular in shape, located on the
front left corner of the compressor compartment and is secured in
place by a mounting bracket with three bolts. See Figure 5-2.
2. Directly above the battery(s) is the battery power switch. Turn the
battery power switch to the Off position.
3. Disconnect the battery connections.
4. Remove the tape securing the battery.
5. Remove the old battery and install the new battery.
6. Reconnect the battery (red to positive and black to negative).
7. Turn the battery power switch to Standby mode ( ).
8. Close lower panel door.
Caution The % of charge can vary depending on the age, usage and
condition of the battery. For a consistent and dependable charge, replace
the battery every 2 years. Replacement batteries must be rechargeable and
are available from Thermo. Refer to the parts list for stock number and
description of the replacement batteries. Dispose of the used batteries in a
safe manner and in accordance with good environmental practices.