Datalogger

Water-cooled
Condenser
Five Inner Door
Thermo Scientific
Dataloggers and ELPRO evaluation software provide monitoring and
documentation of temperature and alarm conditions. The dataloggers have
a memory capacity of 64,000 measured values or data points. Temperature
is measured, stored and displayed. Alarm conditions are recorded.
Optional evaluation software permits data to be downloaded to a PC. A
variety of statistical information is provided through calculations, analysis,
graphs and printed reports. Refer to the ELPRO documentation for
operating instructions for the datalogger.
The water-cooled condenser (P/N 195964 [13cf ], 195965 [17, 23, 28 cf ],
195967 [12, 17, 20 cf chests]) is a factory installed option and requires a
qualified technician at freezer installation. The installation should include
proper adjustment of the regulating valve, which controls the discharge
pressure. Specifications for this option are displayed in Table 6-1.
Table 6-1. Specifications
Water Source
Water Pressure
Water Temperature Range
Inlet Connection
Outlet Connection
Flow Rate Required
Drain Required
The five inner door option (P/N 189405 [13cf ], 189406 [17cf ], 189407
[23cf ] & 195652 [28cf ]) is factory installed. The freezer is converted to
Option
accommodate four adjustable specimen shelves with the fifth "shelf " as the
bottom of the freezer chamber.
Tower
Not to exceed 150 psig
Not to exceed 29.4C (85F)
3.0 gallons (11.4 liters) per minute
No (return line is required)
City
0.5" compression
0.5" compression
1.0 gallon (3.8 liters) per minute
Yes
Model 8600 Series
Section 6
Factory Options
6-9