Install the Temperature
Thermo Scientific
1. Locate the 0.500" pre-punched hole in
the upper left hand back corner of the
Probe
chamber ceiling. Remove the tie wrap
securing the coiled probe/solenoid
harness. Uncoil the probe lead and run
the probe tip (approximately 12")
down through 0.500" porthole (Figure
6-4).
2. As shown in Figure 6-3, thread the
small tie wrap through the openings in
the front of the bracket. Secure the
probe on the back of the bracket with
the tie wrap.
3. Tap #8-32 the two pre-punched holes located on the interior left wall
of the freezer. Mount the bracket. Figure 6-4 shows the back-up probe
mounted on the interior left side wall of the freezer.
Figure 6-4. Mounted
Factory Options
195419
Probe Mounting
Bracket
30037
Small Tie
Wrap
290167
Back-Up Probe
Figure 6-3. Secure Probe
Model 8600 Series
Section 6
6-3