Equipment Overview: Preparation for Use

Connector Identification

WARNING
2-6
A B C
Item
A
B
C
D
E
F
G
H
I
Keep leakage current
within acceptable limits
when connecting
auxiliary equipment to
this device.
M15287-7C
MAC 5000 resting ECG analysis system
2000657-002
D
E
F
Table 2-1. Back Panel Connectors
Name
A
Not functional.
1
Connect a Marquette KISS pump (optional).
2
Connect a local transmission cable, serial line,
Marquette KISS pump, or external modem (optional).
ANA/TTL
Connect a device requiring analog data or TTL
trigger.
EXT.VID.
Connect an external video display.
IR
Point at a MAC 5000 or MUSE CV system's IR
transceiver to transmit or receive ECG data.
card slot
Insert the software card into this slot to run the
system.
ground lug
Connect non-grounded peripheral devices to ensure
equipotential.
mains AC
Insert the mains AC power cable.
power
G
H
I
Description
WARNING
Total system leakage
current must not
exceed 100
microamperes.
MD1325-118A
M15287-9D
Revision B