LeCroy LW110 User Manual page 125

100 ms/s single / dual arbitrary waveform generator
Table of Contents

Advertisement

User Manual LW120
Parameter type
Discrete
TRIGger:SLOPe?
Response and default
The LW120 will return POS, or NEG depending on the present trigger setting. Default value is POS.
TRIGger:SOURce:ADVance{EXTernal|INTernal}
Purpose
This command will set the advance source for the trigger mode.
<EXTernal> will select the external input
<INTernal > will select the internal trigger generator
Parameter type
Discrete
TRIGger:SOURce:ADVance?
Response and default
The LW120 will return EXT, or INT depending on the present trigger source setting. Default value is EXT.
TRIGger:TIMer<interval>
Purpose
This command will set the period for the internal trigger generator.
<interval> will set the timer of the internal trigger generator
Parameter type
Numeric
Parameter range
<interval>
100e-3 to 5e6 in units of seconds
TRIGger:TIMer?
Response and default
The LW120 will return the present internal trigger timer value. The returned value will be in standard
scientific format (for example: 10ms would be returned as 10E-3 – positive numbers are unsigned). Default
value is 1e3.
TRIGger:IMMediate
Purpose
This command will stimulate the LW120 output once. This command will affect the LW120 only after it will
be set to interrupted operation.
*TRG
Purpose
This common command has the same affect as the TRIG:IMM command.