Operation
Overhead Lamp ON/OFF Brightness Controls
The area labeled "LAMP" has 3 separate buttons and a minimum to maximum LED scale. Pressing the
Up or Down Arrow one position at a time will increase or decrease lamp intensity. Pressing the POWER
Switch will apply power to the Overhead Lamp. The LED on the scale will flash indicating power is applied
to the lamp. Pressing the POWER Switch again removes power from the Overhead Lamp and the LED will
remain ON.

Corded Instrument Voltage Selector

Locate the Voltage Limit Switch for corded instruments located on the rear of the stand. Refer to the
Features and Functions section of this manual. Using a small flathead screwdriver, adjust the switch
to the maximum voltage of your corded instrument connected to the Binding Post.
CAUTION: DO NOT SELECT A VOLTAGE THAT EXCEEDS THE VOLTAGE RATING OF THE BULB IN YOUR INSTRUMENT
OR DAMAGE TO THE BULB MAY OCCUR.
Once the maximum voltage has been selected, press the Power Button in the area labeled "VOLTAGE"
to turn your corded instrument on and off. Pressing the Up or Down Arrow one position at a time will
increase or decrease the voltage to the corded instrument connected to the Binding Posts of the stand
within the range of the maximum voltage set by the Voltage Limit Switch on the rear of the stand.

Overhead Lamp

WARNING: DO NOT TOUCH THE LAMP SHADE DIRECTLY NEAR THE BULB OR SERIOUS INJURY MAY OCCUR.
GRASP THE OPEN RING HANDLE AROUND THE LAMP TO ADJUST THE POSITION OF THE LAMP
ASSEMBLY TO AVOID INJURY.
WARNING: TO REDUCE THE RISK OF STATIC SHOCK, DO NOT TOUCH THE OVERHEAD LAMP AND A PATIENT AT
THE SAME TIME.
To operate the Overhead Lamp you must:
1. Switch on the Main ON/OFF Switch on the front of the stand.
2. Press the Lamp POWER Switch on the Control Panel.
3. Press the Left or Right Arrow to increase or decrease lamp intensity.
4. The lamp can also be turned on and off using the rocker switch on the lamp shade.

Chair Operation

CAUTION: THE LIFT MECHANISM IS EQUIPPED WITH A THERMAL CUTOFF SWITCH TO PROTECT THE MECHANISM
FROM OVERHEATING. RAISING AND LOWERING OF THE CHAIR CONTINUOUSLY FOR APPROXIMATELY
ONE AND A HALF MINUTES WILL ACTIVATE THE THERMAL CUTOFF SWITCH AND NOT ALLOW THE
CHAIR TO BE RAISED OR LOWERED. ONCE THE THERMAL CUTOFF SWITCH HAS BEEN ACTIVATED,
SEVERAL MINUTES OF COOLING TIME COMBINED WITH TURNING THE CHAIR MAIN POWER SWITCH
OFF AND THEN ON AGAIN TO RESET THE SYSTEM ARE NECESSARY BEFORE THE LIFT MECHANISM
WILL OPERATE AGAIN. THE TEMPERATURE AT WHICH THE SYSTEM IS ALLOWED TO OPERATE AFTER
COOLING IS MUCH LOWER THAN THE CUTOFF TEMPERATURE, HENCE SEVERAL MINUTES OF
COOLING TIME ARE REQUIRED.
Note: To raise or lower the chair, the Main Power ON/OFF Switch on the front of the stand must be ON.
Endurance Tilt Chair and Stand User's Guide 15250-101 REV-A