Using Tidemaster Control Panel; Pressure Sensor Calibration; Radar Sensor Calibration - Valeport TideMaster Operating Manual

Table of Contents

Advertisement

7.3 Using TideMaster Control Panel

Calibrating the TideMaster using the display panel is essentially the same process as described above.

7.3.1 Pressure Sensor Calibration

Interrupt the TideMaster on start up or during a burst using the left hand menu button.
Navigate to the Pressure Transducer Site Calibration menu
(Menu > Instrument > Press Tdr > Site Cal)
There are three options available.
Enter Low Value
Enter High Value
Calculate Calibration
Use these menu options to follow either the moving or fixed transducer method entering Low/High values
using the appropriate menu option when prompted. When you are happy with the values entered, use the
Calculate Calibration option calculate the calibration factors and update the Gain and Offset parameters.
Run a Cal check. (Menu > Instrument > Press Tdr > Cal Check)
This will display raw data in dBar, and calibrated data in Metres and Feet. The calibrated data should
correspond to the current tide height relative to your datum.

7.3.2 Radar Sensor Calibration

Interrupt the TideMaster on start-up or during a burst using the left hand menu button.
Navigate to the Radar Calibration menu
(Menu > Instrument > Radar > Calibration)
There are three options available:
Datum Offset
Max Limit
Min Limit
© 2021 – Valeport Ltd
TideMaster - Operating Manual: MANUAL-1094978189-1 | issue: 1.1
Page | 36

Advertisement

Table of Contents
loading

Table of Contents