This instrument is an outstanding combination of advanced electronics desin and sensor technology. It’s the most economical choice of a reliable lab benchtop conductivity meter. Please read this maual carefully to properly use and maintain the meter. Apera Instruments reserves the right to update the contents of this manual without giving prior notices.
Calibration 1-4 point Auto Calibration Range 0 to 100˚C; 32 to 212°F Temp. Resolution 0.1˚C; 0.1/1°F Accuracy ±0.5˚C ±1 digit Data Storage 50 groups Numberings, Measurement, Unit, Temperature, Temperature Storage Content Compensation Status Others Power DC9V/300mA IP Ratings IP54 dustproof and spillproof Size and Dimension (240*235*103) mm/1kg Instrument Description...
3.3 Connectors Symbol Connector Name Connector Type COND Socket for conductivity electrode TEMP Socket for temperature sensor DC9V DC9V power socket Φ2.5 direct type 3.4 Stable reading display mode When the measuring value is stable, smiley face icon appears on LCD, see Figure-3.
numberings in RM indicates the number of stored data in current mode (cond/TDS). For example, if 20 groups of data is stored in conductivity mode, and 10 groups in TDS mode, then the numberings for M+ will be “M+30” in both conductivity and TDS mode; the numberings for RM in conductivity mode would be “RM20”, in TDS mode “RM10”.
Table-2 Calibration icon pH standard buffer series 84 μS/cm 1413 μS/cm 4-Point calibration 12.88 mS/cm 111.8 mS/cm 4.3.2 How often to calibrate • We recommend to calibrate the conductivity meter once per month under normal conditions. • If the requirement for accuracy is high or the temperature of the test sample has a big difference to the reference temperature (25 ˚C), we recommend to do the calibration once a week.
4.5 TDS and Conductivity 4.5.1. TDS and conductivity is linear related. The conversion factor is 0.40-1.00. Adjust the factor from parameter P2.6. The factory default setting is 0.71 and please refer to Section 7.4. The meter only needs to be calibrated in Conductivity mode, then after calibration of conductivity, the meter can switch from conductivity to TDS.
P4—Adjust TDS Factor (0.40-1.00) 1. Press , “0.71” flashes, press to adjust from 0.40-1.00; press to confirm. 2. Press to enter P5, or press to return to measurement mode. P5 —(°C—°F) 1. Press , “°C” flashes, press , “°F” flashes; when parameter flashes, press to confirm.
APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS, LLC for a period of two years from the delivery (a six-month limited warranty applies to probes). This warranty does not apply to defects resulting from actions such as misuse (violation of the instructions in this manual or operations in the manner not specified in this manual), improper maintenance, or unauthorized repairs.