Thank you for purchasing the Apera Instruments PH60 Series Premium pH Tester. Please read this manual thoroughly before use in order to properly use the instrument and have a reliable testing experience. Apera Instruments reserves the rights to update the information of this manual without giving prior notices.
1. Battery Installation Please install batteries according to the following steps. Please note direction of batteries: All POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will cause damage to the tester and potential hazards) ﹣ ﹣ ﹣ ﹣ ① Pull the battery cap up ②...
4. Preparation Before Use 4.1 PH60 and PH60F: If it is first-time use or the tester hasn’t been used for over 30 days, pour certain amount (about a quarter of the probe cap) of 3M KCL solution in the probe cap, and soak the probe for about 3 to 15 minutes.
5. pH Calibration 5.1 Rinse the probe in pure water and dry it. 5.2 Pour certain amount (about half volume of calibration bottles) of pH7.00 pH and pH 4.00 buffer solutions in separate calibration bottles. 5.3 Long press to enter calibration mode; Short press to exit.
2 to 3 seconds. 6.2 Special notes a) Applications of each model: Model/Probe Application PH60/Bulb probe Most common liquid measurement such as hydroponics, aquaculture, pools and spas, water treatment, cooling towers, etc. Cheese, sushi rice, meat, fruit, bread, soil, solid culture medium PH60S/Spear probe and semi-solid medium measurement;...
c) PH60F Flat probe testers are mostly for flat surface sample test. ■ For skin test: skin should be without sweat or dirt, nor be overly cleaned (do not use facewash products before testing) to avoid affecting measurement results, dampen skin with some distilled water, slightly force flat probe onto the skin, get readings after value stabilized.
Page 8
7.2 Parameter setup method When turned off, long press to enter parameter setting→Short press to switch P1-P2- …P7→Short parameter flashing→Short press to choose parameter→Short press to confirm→Long press to switch off. 7.3 Parameter setting instruction a) Select standard pH buffer solution (P1): There are two options of standard buffer solutions: USA series and NIST series.
7.00pH, slope is 100%), parameter setting return to initial value. This function can be used when instrument does not work properly in calibration or measurement. Calibrate and measure again after recovering the instrument to factory default status. 8. ORP Measurement Refer to Clause 11 to replace OPR probe (to be purchased separately), press key to enter mode.
11.1 Twist off the probe ring, unplug the probe, plug in new probe (pay attention to probe’s position), and twist on the probe ring. 11.2 The model numbers of replacement probes that are compatible with the tester are: PH60-E (Regular pH glass bulb probe) PH60S-E (Spear pH probe for solids/semi-solids pH testing) ...
We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS, LLC for a period of two years from the delivery (a six-month limited warranty applies to probes).