CEFLA NewTom GiANO HR Technical Manual
Hide thumbs Also See for NewTom GiANO HR:
Table of Contents

Advertisement

Sede legale ed amministrativa / Head Quarter
CEFLA s.c. Via Selice Prov.le 23/a - 40026 Imola (BO) - Italy
tel.+39 0542 653111 / fax +39 0542 653344
After Sales Service
Plant Via Bicocca 14/C 40026 Imola (BO) - Italy
tel.+39 0542 653742 / service.radiology@cefla.it
GiANO HR
T e c h n i c a l m a n u a l
c o d . 9 7 0 7 1 2 3 6 - r e v . 0 2 / 1 0 . 0 6 . 1 9

Advertisement

Table of Contents
loading

Summary of Contents for CEFLA NewTom GiANO HR

  • Page 1 . 9 7 0 7 1 2 3 6 - r e v . 0 2 / 1 0 . 0 6 . 1 9 Sede legale ed amministrativa / Head Quarter After Sales Service CEFLA s.c. Via Selice Prov.le 23/a - 40026 Imola (BO) - Italy Plant Via Bicocca 14/C 40026 Imola (BO) - Italy tel.+39 0542 653111 / fax +39 0542 653344...
  • Page 3 9 7 0 7 1 2 3 6 r e v . 0 2 10.06.2019...
  • Page 5: Table Of Contents

    NewTom GiANO HR - TECHNICAL MANUAL Contents 1. GENERAL WARNINGS ................................3 1.1 Foreword ....................................3 1.2 Safety instructions ................................3 2. PACKAGING AND HANDLING ..............................7 2.1 Handling and storage ................................7 2.2 Damage during shipment ..............................7 3. PREINSTALLATION ..................................9 4. DEVICE INSTALLATION AND CALIBRATION ........................11 4.1 Assembly instructions .................................
  • Page 6 NewTom GiANO HR - TECHNICAL MANUAL 4.4.13 CEPH Verify Mechanical Centering ........................155 4.4.14 Headrest Calibration ............................160 4.4.15 Nasion calibration ...............................163 4.4.16 Backup Calibration Data to File ..........................165 4.4.17 Aligning the ear guide rings ..........................167 4.5 3D Calibration ..................................171 4.5.1 Beam Limiter Test ..............................171 4.5.2 Blank Acquisition and Daily Check ........................178...
  • Page 7: General Warnings

    These technical instructions are addressed to the personnel in charge of repair and/or maintenance operations of CEFLA s.c. equipment and contain all the necessary information. CEFLA s.c. shall be responsible for the safety, reliability and efficiency of the equipment provided that: • installation,  any  modifications,  settings  or  repairs  are  made  by  authorised  technical  personnel  using  CEFLA  s.c.  original spare parts • In case of installation in medical locations: the electrical installation of the relevant location complies with IEC 60364- 7-710:2002 Standards (Standards on electrical installations of medical locations) or with the equivalent Standards in ...
  • Page 8 NewTom GiANO HR - TECHNICAL MANUAL SYMBOL DESCRIPTION ATTENTION: Refer to instruction manual-booklet ATTENTION: Disconnect mains plug from electrical outlet WARNING: See the technical manual WARNING: Failure to observe instructions may result in equipment damage or injury to the user WARNING: High voltage WARNING: Crushing of hands...
  • Page 9 NewTom GiANO HR - TECHNICAL MANUAL SYMBOL DESCRIPTION WARNING: Combustible and flammable materials DANGER: Tipping over DANGER: Heavy object. Two persons are required WARNING! ALWAYS DISCONNECT power supply before performing any maintenance or cleaning operation. The packed device must be handled using, where possible, suitable mechanical means (forklift, pallet truck, etc.) and following the indications on the package.
  • Page 10 NewTom GiANO HR - TECHNICAL MANUAL GENERAL WARNINGS...
  • Page 11: Packaging And Handling

    • if the goods are delivered “carriage forward”, the receiver shall notify the shipping agent • if the goods are shipped “carriage free” or “free delivered”, the consigner shall notify the shipping agent and file a  claim for damages. In this case, CEFLA s.c. shall be informed as soon as possible. The damaged parts returned to CEFLA s.c. for replacement shall be placed inside the same damaged package. PACKAGING AND HANDLING...
  • Page 12 NewTom GiANO HR - TECHNICAL MANUAL Return to CEFLA s.c. shall be “carriage free” (at the expense of the consigner), while shipment back to the customershall  be at the expense of CEFLA s.c. (carriage free). PACKAGING AND HANDLING...
  • Page 13: Preinstallation

    NewTom GiANO HR - TECHNICAL MANUAL 3. PREINSTALLATION Follow instructions described in the supplied INSTALLATION PLAN. PREINSTALLATION...
  • Page 14 NewTom GiANO HR - TECHNICAL MANUAL PREINSTALLATION...
  • Page 15: Device Installation And Calibration

    NewTom GiANO HR - TECHNICAL MANUAL 4. DEVICE INSTALLATION AND CALIBRATION 4.1 ASSEMBLY INSTRUCTIONS Installation of the device unit must be done by qualified technicians in accordance with the mechanical and electrical  assembly instructions defined as follows. Check that the voltage indicated on the system info plate corresponds to that  of the electrical system. The following instruments are necessary to perform all the procedure properly: Allen wrench set Screwdrivers Scissors Spirit level Electric cable Threaded bars or bolts Network cables RJ12 telephone cable Calibration kit 4.1.1 REMOVAL OF PACKAGING...
  • Page 16: Transport With Cart

    NewTom GiANO HR - TECHNICAL MANUAL 4.1.2 TRANSPORT WITH CART Screw on the feet (if the device is installed without the base);  secure  the  cart  to  the  base  of  the  device  with  the  screws provided. Place the cart’s brackets on top of the post’s metal base.
  • Page 17: Support Base Floor Mounting Instructions

    NewTom GiANO HR - TECHNICAL MANUAL 4.1.4 SUPPORT BASE FLOOR MOUNTING INSTRUCTIONS With the aid of a fork-lift, position the device above the support base; lower the  device in line with the fixing holes; Fix the device to the support base using the five supplied bolts; -  Remove the fork-lift; Check for proper levelling with a spirit level and if necessary adjust the feet.
  • Page 18: Patient Arm Installation

    NewTom GiANO HR - TECHNICAL MANUAL 4.1.7 PATIENT ARM INSTALLATION DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 19 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 20 NewTom GiANO HR - TECHNICAL MANUAL Mainboard DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 21: Electrical Connecions On Mainboard

    NewTom GiANO HR - TECHNICAL MANUAL COL+ COLT COLS COL- GND_OUT EXP_OK+ USER_ENB+ EX P_OK- GND_OUT VCC_OUT USER_SYNC+ USER_ENB- USER_SYNC- 4.1.8 ELECTRICAL CONNECIONS ON MAINBOARD GND_OUT EX P_OK+ USER_ENB+ EX P_OK- Mainboard GND_OUT Console power VCC_OUT USER_SYNC+ USER_ENB- USER_SYNC- supply board...
  • Page 22: Electrical Connections On The Pfc Board

    NewTom GiANO HR - TECHNICAL MANUAL 4.1.9 ELECTRICAL CONNECTIONS ON THE PFC BOARD MAIN SUPPLY INTERLOCK EMERGENCY STOP X-RAY BUTTON (1: VCC / 2: BUTTON / 3: READY / 4: X-RAY) READY LAMP (1-2) X-RAY ON LAMP (3-4) COMPULSORY OPTIONAL For further information, see the supplied INSTALLATION DIAGRAM.
  • Page 23: Ethernet Connections And X-Ray Emission Key

    NewTom GiANO HR - TECHNICAL MANUAL 4.1.10 ETHERNET CONNECTIONS AND X-RAY EMISSION KEY CONNECTOR FUNCTION BROWN X-RAY Emission WHITE Green “Ready” LED BLUE Orange “X-Ray” LED For further information, see the supplied INSTALLATION DIAGRAM. DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 24: Ceph Arm Installation

    NewTom GiANO HR - TECHNICAL MANUAL 4.1.11 CEPH ARM INSTALLATION DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 25 NewTom GiANO HR - TECHNICAL MANUAL RIGHT LEFT DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 26 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 27 NewTom GiANO HR - TECHNICAL MANUAL Mainboard DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 28 NewTom GiANO HR - TECHNICAL MANUAL CLICK! DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 29: Electrical Connecions On Mainboard

    NewTom GiANO HR - TECHNICAL MANUAL 4.1.12 ELECTRICAL CONNECIONS ON MAINBOARD K42/47 CEPH KEYBOARD CEPH MOTOR COL+ COLT COLS COL- GND_OUT EXP_OK+ USER_ENB+ EX P_OK- GND_OUT VCC_OUT USER_SYNC+ USER_ENB- USER_SYNC- GND_OUT EX P_OK+ USER_ENB+ EX P_OK- GND_OUT VCC_OUT USER_SYNC+ USER_ENB-...
  • Page 30: Patient Arm's Handles Assembly

    NewTom GiANO HR - TECHNICAL MANUAL 4.1.13 PATIENT ARM’S HANDLES ASSEMBLY DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 31: Check During The Assembling

    NewTom GiANO HR - TECHNICAL MANUAL 4.1.14 CHECK DURING THE ASSEMBLING PREINSTALLATION CHECK RESULT OK Intact external package  Check compliance of delivery documents  Check ID label/package labels match  Check place of installation and supplies*  POWER SUPPLY CHECK Connection of power and earth leads ...
  • Page 32 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Assembly instructions...
  • Page 33: Cover Installation

    NewTom GiANO HR - TECHNICAL MANUAL 4.2 COVER INSTALLATION Screw kit SYMBOL DESCRIPTION T/BOMB M4x8 T/BOMB M4x14 TCEI M4x12 WASHER D 4.3 x9 TPSEI M4x16 CLIP M4 MAX 2.5 mm TCEI SCREW M4X20 STAINLESS DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 34 NewTom GiANO HR - TECHNICAL MANUAL Cover assembly tasks must be carried out in the order illustrated in the drawing above.   The screws C must only be placed in position: they must only be tightened after positioning the cover on the  unit so these screws line up with the slots highlighted in the subsequent diagrams (in the second image they are not visible but they are analogous to those in the first image).
  • Page 35 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 36 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 37 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 38 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 39 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 40 NewTom GiANO HR - TECHNICAL MANUAL CLICK! DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 41 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 42 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 43 NewTom GiANO HR - TECHNICAL MANUAL CLICK! DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 44 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » Cover installation...
  • Page 45: Pc Configuration

    NewTom GiANO HR - TECHNICAL MANUAL 4.3 PC CONFIGURATION 4.3.1 SYSTEM REQUIREMENTS The device works with 2 Ethernet connections, one for generic communication with the PC (that can be connected to a local area network) and another one (which must absolutely be point-to-point) dedicated to image acquisition from the  3D panel, therefore connecting the device to the INTEL GIGABIT network card of the computer.
  • Page 46 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 47: Power Management Settings

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.3.1 POWER MANAGEMENT SETTINGS DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 48 NewTom GiANO HR - TECHNICAL MANUAL Disabling “Fast start-up” DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 49: Operating System Optimization Settings

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.3.2 OPERATING SYSTEM OPTIMIZATION SETTINGS DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 50 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 51 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 52: Disabling "Windows Auto-Update Service

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.3.3 DISABLING “WINDOWS AUTO-UPDATE SERVICE” DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 53 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 54: Disabling Windows Update Sharing Over The Network (Low Data Traffic)

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.3.4 DISABLING WINDOWS UPDATE SHARING OVER THE NETWORK (LOW DATA TRAFFIC) DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 55: Firewall Settings

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.3.5 FIREWALL SETTINGS Firewall software programs (such as for example the one integrated into Windows) limit the network traffic between  the PC connected to the device and the local network. We recommend to disable these firewall programs to prevent  any problem in communicating with the device, downloading images and sharing software licences over the network. In case they cannot be disabled, it is nevertheless necessary to add the following exceptions: UDP PORT 12232: always allow incoming/outgoing traffic from the LAN for all the programs UDP PORT 12233: always allow incoming/outgoing traffic from the LAN for all the programs UDP PORT 12234: always allow incoming/outgoing traffic from the LAN for all the programs TCP PORT 12235: always allow incoming/outgoing traffic from the LAN for all the programs TCP PORT 12236: always allow incoming/outgoing traffic from the LAN for all the programs TCP/UDP PORT 1947: always allow incoming/outgoing traffic from the LAN for all the programs TCP PORT 13234: always allow incoming/outgoing traffic from the LAN for all the programs TCP PORT 5790: always allow incoming/outgoing traffic from the LAN for all the programs 4.3.3.6 “WINDOWS DEFENDER” SETTINGS DEVICE INSTALLATION AND CALIBRATION »...
  • Page 56 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 57 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 58: Device Installation Settings

    NewTom GiANO HR - TECHNICAL MANUAL AFTER SOFTWARE INSTALLATION (see paragraphs 4.3.6, 4.3.11) 4.3.3.7 DEVICE INSTALLATION SETTINGS DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 59 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 60 NewTom GiANO HR - TECHNICAL MANUAL gpedit.msc DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 61 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 62: Pc-Device Connection

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.4 PC-DEVICE CONNECTION The device may be connected to a local area network or directly to the main workstation. In both cases, the Ethernet connection for sensors must be point-to-point and is always managed autonomously by the software and the machine.
  • Page 63 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 64: Case 2 - Device-Pc Connection Over The Network (Setting Adynamic Ip Address On The Network Card)

    NewTom GiANO HR - TECHNICAL MANUAL 1->254 160 (IP Device) 161 (IP Console) 4.3.4.2 CASE 2 - DEVICE-PC CONNECTION OVER THE NETWORK (SETTING A DYNAMIC IP ADDRESS ON THE NETWORK CARD) It is necessary to set a dynamic IP address for the network card dedicated to PC – device communication.
  • Page 65 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 66: Intel Gigabit" Network Card Configuration

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.5 “INTEL GIGABIT” NETWORK CARD CONFIGURATION DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 67 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 68 NewTom GiANO HR - TECHNICAL MANUAL WARNING: maybe the “Receive Buffers” item is not displayed directly in this window. If it is not displayed, search for the “Performance Options” item and click on “Properties”. The “Receive Buffers” value is displayed there. WARNING: maybe the “Interrupt Moderation”...
  • Page 69 NewTom GiANO HR - TECHNICAL MANUAL WARNING: maybe the “Interrupt Moderation Rate” item is not displayed directly in this window. If it is not displayed, search for the “Performance Options” item and click on “Properties”. The “Interrupt Moderation Rate” value is displayed there.
  • Page 70 NewTom GiANO HR - TECHNICAL MANUAL Avoid using 192.168.2.xxx subnet for other network adapters in the PC DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 71: Acquisition Server Plus" Software Installation

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.6 “ACQUISITION SERVER PLUS” SOFTWARE INSTALLATION Software Acquisition Server Plus Acquisition Server Plus DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 72 NewTom GiANO HR - TECHNICAL MANUAL Confirm as requested by the installer to complete the installation. WARNING: Drivers for communication with 2D sensors will be installed during this installation. DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 73: Acquisition Server Plus" Configuration

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.7 “ACQUISITION SERVER PLUS” CONFIGURATION USER MODE: DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 74 NewTom GiANO HR - TECHNICAL MANUAL cefla DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 75: Device Search And Connection

    NewTom GiANO HR - TECHNICAL MANUAL SERVICE MODE: 4.3.7.1 DEVICE SEARCH AND CONNECTION DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 76 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 77 NewTom GiANO HR - TECHNICAL MANUAL NOTE: should it be necessary to change the device IP address for any need, it is possible to change it from the menu “Settings -> Device Options” A fixed IP can be set in this window in the field “Fall back Ip address”. Removing the tick from  “DHCP” will make it impossible for the device to obtain a dynamic address by a router.
  • Page 78: Sensor Search And Connection

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.7.2 SENSOR SEARCH AND CONNECTION DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 79 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 80 NewTom GiANO HR - TECHNICAL MANUAL WARNING IN CASE OF CEPH SENSORS: In case of a configuration with a mobile sensor, move it from “PAN” to “CEPH”  position. DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 81 NewTom GiANO HR - TECHNICAL MANUAL Under normal operating conditions, the lower box may be red: this does NOT indicate a connection problem since PAN/CEPH sensors are usually off. They are on only when they are being used. DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 82: Cbct Panel Installation (Only For 3D)

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.8 CBCT PANEL INSTALLATION (ONLY FOR 3D) WARNING: each panel is equipped with a specific installation CD, containing data that are not interchangeable with other units. Carefully keep the CD and make sure that at least one backup copy is available. Run and complete the installation leaving the default settings.
  • Page 83 NewTom GiANO HR - TECHNICAL MANUAL cefla DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 84 NewTom GiANO HR - TECHNICAL MANUAL CBCT Panel ON DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 85 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 86 NewTom GiANO HR - TECHNICAL MANUAL The installer will check whether a previous version of Pleora Driver is already present and will uninstall it if necessary. WARNING: this procedure may take a long time, do not interrupt the process. During the procedure the PC will restart and resume installation.
  • Page 87: Driver Installation Checks

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.10 DRIVER INSTALLATION CHECKS DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 88 NewTom GiANO HR - TECHNICAL MANUAL WARNING: perform the procedures (1 -> 5) for all the network cards that may be present on the PC, except for  the one dedicated to the communication with sensors (INTEL GIGABIT). DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 89: Nnt Software Installation And Configuration

    NewTom GiANO HR - TECHNICAL MANUAL If items (8 -> 9) are missing, this indicates that the driver installation was not successful and it must be repeated. Missing item 8 (only in case of 3D!): see chapter “Pleora driver intallation” Missing item 9: see chapter “Acquisition Server Plus software installation“ 4.3.11 NNT SOFTWARE INSTALLATION AND CONFIGURATION Software NNT Software NNT Software The software used to manage the images downloaded from the device is NNT, installed in the version MAIN WORKSTATION in case of 3D device, or in the version NNT Station in case of 2D device only.
  • Page 90 NewTom GiANO HR - TECHNICAL MANUAL 2D ONLY 3D ONLY DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 91 NewTom GiANO HR - TECHNICAL MANUAL Confirm as requested by the installer to complete the installation. DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 92 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 93 NewTom GiANO HR - TECHNICAL MANUAL Enter Service Mode: METHOD 1 DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 94 NewTom GiANO HR - TECHNICAL MANUAL METHOD 2 306090 3D ONLY 336699 DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 95 NewTom GiANO HR - TECHNICAL MANUAL 2D - 3D 12345 DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 96 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 97 NewTom GiANO HR - TECHNICAL MANUAL 3D ONLY DEVICE CONNECTED DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 98 NewTom GiANO HR - TECHNICAL MANUAL Under normal operating conditions, the lower box may be red: this does NOT indicate a connection problem since PAN/CEPH sensors are usually off. They are on only when they are being used. DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 99: Onboard Console Connection

    NewTom GiANO HR - TECHNICAL MANUAL 4.3.12 ONBOARD CONSOLE CONNECTION DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 100 NewTom GiANO HR - TECHNICAL MANUAL 192.168.1.160 PC - HP YOUR DEVICE 192.168.1.160 PC - HP YOUR DEVICE DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 101 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 102 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » PC configuration...
  • Page 103: Calibration

    NewTom GiANO HR - TECHNICAL MANUAL 4.4 2D CALIBRATION REQUIRED INSTRUMENTS Calibration kit, including: • Support plate • Dentition phantom • Aluminium filter • Laser reference sheet • Allen wrench set • Screwdriver set • Pliers 4.4.1 TUBE HEAD WARM-UP PERFORM THIS OPERATION ONLY IF THE DEVICE REMAINS UNUSED FOR MORE...
  • Page 104 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 105 NewTom GiANO HR - TECHNICAL MANUAL Shot Parameters 1000 Time exposure (ms) Time (ms) Repetitions 1000 1000 1000 1000 1000 1000 5000 5000 3D ONLY Perform Daily Check (see paragraph 4.5.2) DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 106: Preliminary Actions

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.2 PRELIMINARY ACTIONS DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 107 NewTom GiANO HR - TECHNICAL MANUAL cefla DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 108: Machine Configuration

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.3 MACHINE CONFIGURATION DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 109 NewTom GiANO HR - TECHNICAL MANUAL / 2D Model CHECK / ABSENT Device Serial No. = Device Name Just 1 sensor = NULL NO CBCT = NULL NO CBCT = NULL √ √ always always √ √ always only 3D device √...
  • Page 110: Column Calibration

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.4 COLUMN CALIBRATION DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 111 NewTom GiANO HR - TECHNICAL MANUAL CLICK! BITEWING POSITION 1015 min. DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 112 NewTom GiANO HR - TECHNICAL MANUAL CLICK! BITEWING POSITION DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 113 NewTom GiANO HR - TECHNICAL MANUAL Set SW limits: these values add a further tolerance to the limit switch: it is recommended to remove 10 mm from the  upper limit and add 10 mm to the lower one. DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 114: Pan Verify Xray Alignement

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.5 PAN VERIFY XRAY ALIGNEMENT TA R G E T DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 115 NewTom GiANO HR - TECHNICAL MANUAL MOTORISED HORIZONTAL COLLIMATOR: with the first acquisitions, the procedure searches for sensor sides and  asks to evaluate the correct rotation; then it refines lower and upper collimation.  NON-MOTORISED HORIZONTAL COLLIMATOR: the horizontal adjustment, if necessary, must be performed as in paragraph “Horizontal Adjustment”. DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 116 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 117 NewTom GiANO HR - TECHNICAL MANUAL If one of these situations occurs, it is necessary to manually rotate the collimator (See paragraph “Manual rotation of the collimator”). DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 118 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 119 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 120 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 121 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 122: Collimator Manual Rotation

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.5.1 COLLIMATOR MANUAL ROTATION (TO BE PERFORMED WHEN STRICTLY NECESSARY) DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 123 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 124 NewTom GiANO HR - TECHNICAL MANUAL TARGET DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 125: Horizontal Adjustment

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.5.2 HORIZONTAL ADJUSTMENT 2D ONLY DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 126 NewTom GiANO HR - TECHNICAL MANUAL TA R G E T DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 127: Pan Verify Sensor Compliance

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.6 PAN VERIFY SENSOR COMPLIANCE DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 128 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 129: Pan Verify Sensor Calibration

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.7 PAN VERIFY SENSOR CALIBRATION DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 130 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 131 NewTom GiANO HR - TECHNICAL MANUAL 12’ DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 132: Pan Verify Mechanical Centering

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.8 PAN VERIFY MECHANICAL CENTERING DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 133 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 134 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 135 NewTom GiANO HR - TECHNICAL MANUAL CHECK! » 4.1.7 DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 136: Laser Calibration

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.9 LASER CALIBRATION DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 137 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 138 NewTom GiANO HR - TECHNICAL MANUAL If the laser turns off, any of the movement buttons of the column can be pressed to turn it on again. DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 139 NewTom GiANO HR - TECHNICAL MANUAL TA R G E T DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 140: Ceph Verify X-Ray Alignment

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.10 CEPH VERIFY X-RAY ALIGNMENT (ONLY IF CEPH ARM IS PRESENT) TA R G E T DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 141 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 142 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 143 NewTom GiANO HR - TECHNICAL MANUAL If one of these situations occurs, it is necessary to manually rotate the collimator (See paragraph “Manual rotation of the collimator”). DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 144 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 145 NewTom GiANO HR - TECHNICAL MANUAL If the image is not vertically centred, follow the instructions in paragraph “Manual adjustment of vertical position” DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 146 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 147 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 148: Manual Rotation

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.10.1 MANUAL ROTATION (TO BE PERFORMED ONLY IF INDICATED IN THE CALIBRATION PHASE) Ø 2,5mm DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 149 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 150 NewTom GiANO HR - TECHNICAL MANUAL TA R G E T DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 151: Manual Adjustment Of Vertical Position

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.10.2 MANUAL ADJUSTMENT OF VERTICAL POSITION DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 152 NewTom GiANO HR - TECHNICAL MANUAL TA R G E T DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 153: Ceph Verify Sensor Compliance

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.11 CEPH VERIFY SENSOR COMPLIANCE (ONLY IF CEPH ARM IS PRESENT) DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 154 NewTom GiANO HR - TECHNICAL MANUAL Before start procedure remove any object between XRay source and sensor DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 155 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 156: Ceph Verify Sensor Calibration

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.12 CEPH VERIFY SENSOR CALIBRATION (ONLY IF CEPH ARM IS PRESENT) DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 157 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 158 NewTom GiANO HR - TECHNICAL MANUAL 12’ DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 159: Ceph Verify Mechanical Centering

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.13 CEPH VERIFY MECHANICAL CENTERING (ONLY IF CEPH ARM IS PRESENT) DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 160 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 161 NewTom GiANO HR - TECHNICAL MANUAL TA R G E T » 4.1.11 CHECK! DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 162 NewTom GiANO HR - TECHNICAL MANUAL If the image is not horizontally centred, follow the instructions to manually adjust the settings 1.95 1.95 TARGET DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 163 NewTom GiANO HR - TECHNICAL MANUAL 1.75 1.75 TARGET DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 164: Headrest Calibration

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.14 HEADREST CALIBRATION (ONLY IF CEPH ARM IS PRESENT) DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 165 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 166 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 167: Nasion Calibration

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.15 NASION CALIBRATION (ONLY IF CEPH ARM IS PRESENT) DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 168 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 169: Backup Calibration Data To File

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.16 BACKUP CALIBRATION DATA TO FILE DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 170 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 171: Aligning The Ear Guide Rings

    NewTom GiANO HR - TECHNICAL MANUAL 4.4.17 ALIGNING THE EAR GUIDE RINGS To check the correct alignment it is possible to take the image by using a CEPH mechanical positioning window or by launching the virtual keyboard and performing a CEPH LL examination.
  • Page 172 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 173 NewTom GiANO HR - TECHNICAL MANUAL TARGET VERTICAL ALIGNMENT DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 174 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 2D calibration...
  • Page 175: Calibration

    NewTom GiANO HR - TECHNICAL MANUAL 4.5 3D CALIBRATION Required instruments 3D calibration kit, including: • cylinder with two rows of metal spheres. • support plate All covers must be assembled. 4.5.1 BEAM LIMITER TEST DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 176 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 177 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 178 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 179 NewTom GiANO HR - TECHNICAL MANUAL TA R G E T Left 10,5 Right Left 10,5 Right Left 10,5 Right Positioning Down Down Down [13 x 8] [13 x 8] [13 x 8] DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 180 NewTom GiANO HR - TECHNICAL MANUAL ±0,1 ±0,5 15,7 15,5 DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 181 NewTom GiANO HR - TECHNICAL MANUAL 10,5 Left Right Down DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 182: Blank Acquisition And Daily Check

    NewTom GiANO HR - TECHNICAL MANUAL When passing from one FOV to the next, remove the tick from “Enable Manual Tuning” to allow the collimator to reposition itself according to the chosen FOV. 4.5.2 BLANK ACQUISITION AND DAILY CHECK IMPORTANT: If requested by the customer, select menu “Tools-->Force always use dead man button for automated procedure”...
  • Page 183 NewTom GiANO HR - TECHNICAL MANUAL TEST 1 Set up device for CBCT / 3D exam TEST 2 Detector initialization TEST 3 X-Ray source calibration TEST 4 Verification of stability - consistency of emitted/measured dose TEST 5 Verification of consistency of images acquired at different frame rates TEST 6 Verification of the background image (without RX) of detector Test Status Running... Test n° 1 Test n°...
  • Page 184 NewTom GiANO HR - TECHNICAL MANUAL Test Status Test n° 1 Completed Test n° 2 Completed Test n° 3 Running... Test n° 4 Test n° 5 Test n° 6 Test Status Test n° 1 Completed Test n° 2 Completed Test n° 3...
  • Page 185 NewTom GiANO HR - TECHNICAL MANUAL TARGET DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 186: Cylindrical Test Phantom Acquisition ("Mattarellum")

    NewTom GiANO HR - TECHNICAL MANUAL 4.5.3 CYLINDRICAL TEST PHANTOM ACQUISITION (“MATTARELLUM”) DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 187 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 188 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 189 NewTom GiANO HR - TECHNICAL MANUAL The software has a certain tolerance: even if the spheres are not inside the respective positions, the important thing is that the Mattarellum does not generate a software error and that it is therefore possible to proceed with calibration.
  • Page 190 NewTom GiANO HR - TECHNICAL MANUAL TARGET DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 191 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 192 NewTom GiANO HR - TECHNICAL MANUAL Depending on the device model, 1, 2 or 3 acquisitions will be performed. 0.26 1.82 DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 193: Calibration Backup

    NewTom GiANO HR - TECHNICAL MANUAL CLOSE TARGET 4.5.4 3D CALIBRATION BACKUP DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 194 NewTom GiANO HR - TECHNICAL MANUAL Backup 3D DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 195 NewTom GiANO HR - TECHNICAL MANUAL RESTORE Backup 3D Software DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 196: Qa Phantom Scan (Optional)

    NewTom GiANO HR - TECHNICAL MANUAL 4.5.5 QA PHANTOM SCAN (OPTIONAL) Check for evaluation of 3D images quality. DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 197 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 198 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 199 NewTom GiANO HR - TECHNICAL MANUAL TARGET DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 200 NewTom GiANO HR - TECHNICAL MANUAL Ctrl Ctrl DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 201 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 202 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 203 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 204 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 205 NewTom GiANO HR - TECHNICAL MANUAL 27.20 0.47 DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 206 NewTom GiANO HR - TECHNICAL MANUAL CLOSE TARGET DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 207 NewTom GiANO HR - TECHNICAL MANUAL The number of available FOV will vary from 2 to 6, depending on the device model. DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 208 NewTom GiANO HR - TECHNICAL MANUAL DEVICE INSTALLATION AND CALIBRATION » 3D calibration...
  • Page 209: Control Panels And Display Symbols

    NewTom GiANO HR - TECHNICAL MANUAL 5. CONTROL PANELS AND DISPLAY SYMBOLS 5.1 CONSOLE ONBOARD THE MACHINE VERTICAL LASER POSITIONING (FORWARD) VERTICAL LASER POSITIONING (BACKWARD) for 2D projections for 2D projections CRANIOSTAT MOVEMENT (UPWARD) CRANIOSTAT MOVEMENT (DOWNWARD) for 3D projections...
  • Page 210: Pushbutton Panel On Ceph Arm

    NewTom GiANO HR - TECHNICAL MANUAL 5.2 PUSHBUTTON PANEL ON CEPH ARM COLUMN UPWARD MOVEMENT STANDBY COLUMN DOWNWARD MOVEMENT X-RAY READY INTERLOCK CONTROL PANELS AND DISPLAY SYMBOLS...
  • Page 211: Service Menu

    NewTom GiANO HR - TECHNICAL MANUAL 6. SERVICE MENU Method 1 SERVICE MENU...
  • Page 212 NewTom GiANO HR - TECHNICAL MANUAL Method 2 306090 SERVICE MENU...
  • Page 213: Circuit Boards

    NewTom GiANO HR - TECHNICAL MANUAL 7. CIRCUIT BOARDS IMPORTANT:  Before  carrying  out  any  maintenance  work  on  electronic  boards  or  wiring  it  is  necessary to switch off the machine and wait until all the LEDs on the boards go out. Do not  disconnect/reconnect wires or leads with the machine on. 7.1 UNIT BLOCK DIAGRAM CIRCUIT BOARDS...
  • Page 214: Mainboard (97661808)

    NewTom GiANO HR - TECHNICAL MANUAL 7.2 MAINBOARD (97661808) Mainboard CIRCUIT BOARDS...
  • Page 215 NewTom GiANO HR - TECHNICAL MANUAL SCH. CONVERTER 97661807 K35 97522199 CEPH TERMINAL BLOCK BACKLIGHT LOGO 97522483 97661913 K1-K11 97522217 97661539 FC-Y FC-R FC-X 97522266 97522268 97522270 ASSE X SCH. INVERTER 97661954 K3 97522267 97522432 POTENZ. COL+ COLT COLS 97522283 COL- SCH.
  • Page 216: Connector List

    NewTom GiANO HR - TECHNICAL MANUAL 7.2.1 CONNECTOR LIST CONNECTOR DESCRIPTION 12/48V power supply towards CONVERTER X axis motor PAN Synchronization signal CAN for converter X limit switch Y axis motor Y limit switch LAN towards workstation CAN for Inverter board...
  • Page 217 NewTom GiANO HR - TECHNICAL MANUAL DL16 Yellow XRAY On during emission DL18 Yellow Flashing if LAN is broadcasting DL19 Green LINK Flashing if LAN is connected DL22 Green 3.3V ON if 3.3V power supply present DL23 Green ON if 24V power supply present...
  • Page 218: Pfc (97661806)

    NewTom GiANO HR - TECHNICAL MANUAL 7.3 PFC (97661806) CIRCUIT BOARDS...
  • Page 219 NewTom GiANO HR - TECHNICAL MANUAL CIRCUIT BOARDS...
  • Page 220: Connector List

    NewTom GiANO HR - TECHNICAL MANUAL 7.3.1 CONNECTOR LIST CONNECTOR DESCRIPTION 395V power supply connection K2/K3 Main switch Single-phase 115-240 Vac 50/60 Hz general power supply Connector for X-ray command External interlock input connection Connector for External Emergency button External lamps connection (pin 1-2 ready, 3-4 emission)
  • Page 221: Inverter (97661954)

    NewTom GiANO HR - TECHNICAL MANUAL 7.4 INVERTER (97661954) Inverter CIRCUIT BOARDS...
  • Page 222 NewTom GiANO HR - TECHNICAL MANUAL POT_VCC POT_TRIM POT_GND CIRCUIT BOARDS...
  • Page 223: Connector List

    NewTom GiANO HR - TECHNICAL MANUAL 7.4.1 CONNECTOR LIST CONNECTOR DESCRIPTION CAN Connettor towards PFC CAN Connettor towards Main Board 24V towards console power supply 24V power supply towards MAIN Motor Hall sensor connector Motor power supply connector Motor thermal connector...
  • Page 224: Converter (97661807)

    NewTom GiANO HR - TECHNICAL MANUAL 7.5 CONVERTER (97661807) Converter CIRCUIT BOARDS...
  • Page 225 NewTom GiANO HR - TECHNICAL MANUAL SCHEDA MAIN 97522199 K1 3D INTERFACE 97661579 SCHEDA INVERTER 97661954 X2 97522192 STEPPER MOT. A (COLLIMAT.) 97522277 STEPPER MOT. U (COLLIMAT.) 97522279 STEPPER MOT. V (REVOLVER) 97522281 CANH SCHEDA MAIN 97522199 K4 CANS 97522198...
  • Page 226: Connector List

    NewTom GiANO HR - TECHNICAL MANUAL 7.5.1 CONNECTOR LIST CONNECTOR DESCRIPTION Primary horizontal collimator motor 3D panel Synchronization signal Horizontal collimator limit switch Primary vertical collimator motor Vertical collimator limit switch Revolver motor K15/K17 Revolver limit switch Monoblock High voltage Monoblock filament power supply...
  • Page 227: Console Power Supply (97662070)

    NewTom GiANO HR - TECHNICAL MANUAL DL65 Green ON if driver power supply present DL66 Green +5VA On if +5V power supply present DL68 Green ON if driver power supply present DL69 Green VCC CAN ON if isolated 5V present 7.6 CONSOLE POWER SUPPLY (97662070)
  • Page 228 NewTom GiANO HR - TECHNICAL MANUAL CIRCUIT BOARDS...
  • Page 229: Connector List

    NewTom GiANO HR - TECHNICAL MANUAL 7.6.1 CONNECTOR LIST CONNECTOR DESCRIPTION Console board power supply 24V from inverter board LAN switch power supply 7.6.2 DIAGNOSTIC LEDS NAME COLOUR SCREEN PRINTING DESCRIPTION Green SWITCH ON if LAN switch power supply present...
  • Page 230: Connector List

    NewTom GiANO HR - TECHNICAL MANUAL 7.7.1 CONNECTOR LIST CONNECTOR DESCRIPTION K1/K11 MAIN board connection Ear guides potentiometer Nasion potentiometer Cephalostat drum sensors CIRCUIT BOARDS...
  • Page 231: Console (97662111)

    NewTom GiANO HR - TECHNICAL MANUAL 7.8 CONSOLE (97662111) Console CIRCUIT BOARDS...
  • Page 232: Connector List

    NewTom GiANO HR - TECHNICAL MANUAL 7.8.1 CONNECTOR LIST CONNECTOR DESCRIPTION CN2 (LAN) LAN towards switch Gigabit CN14 (POWER) Power supply from K1 on console power supply board CIRCUIT BOARDS...
  • Page 233: Sensor Interface (97661579)

    NewTom GiANO HR - TECHNICAL MANUAL 7.9 SENSOR INTERFACE (97661579) Sensor interface CIRCUIT BOARDS...
  • Page 234: Connector List

    NewTom GiANO HR - TECHNICAL MANUAL CONVERTER K3 3D PANEL 3D PANEL 7.9.1 CONNECTOR LIST CONNECTOR DESCRIPTION Sensor synchronisation / supply from CONVERTER 24V DC Panel supply Sensor synchronisation 7.9.2 DIAGNOSTIC LEDS NAME COLOUR SCREEN PRINTING DESCRIPTION Green 48V from CONVERTER board...
  • Page 235: 5-Port Switch

    NewTom GiANO HR - TECHNICAL MANUAL 7.10 5-PORT SWITCH 5-Port Switch CIRCUIT BOARDS...
  • Page 236 NewTom GiANO HR - TECHNICAL MANUAL PAN SENSOR (GLAN4) CEPH SENSOR (GLAN5) 3D PANEL (GLAN3) (SWITCH GIGABIT) from K30 - MAINBOARD PC - INTEL GIGABIT (GLAN1) MAIN (LAN4) CONSOLE (GLAN3) (SWITCH LAN) from K3 - CONSOLE POWER SUPPLY BOARD PC - COMMUNICATION (LAN)
  • Page 237: Connectivity

    NewTom GiANO HR - TECHNICAL MANUAL 8. CONNECTIVITY 8.1 CONSOLE IP ADDRESS AMENDMENT SELECT EDIT SERVICE MENU...
  • Page 238: Sensors / 3D Panel Ip Address Amendment

    NewTom GiANO HR - TECHNICAL MANUAL 8.2 2D SENSORS / 3D PANEL IP ADDRESS AMENDMENT CONTACT SERVICE 8.3 FIRMWARE UPGRADE UPDATE UPDATE SERVICE MENU...
  • Page 239 NewTom GiANO HR - TECHNICAL MANUAL SERVICE MENU...
  • Page 240: Remote Support

    NewTom GiANO HR - TECHNICAL MANUAL 8.4 REMOTE SUPPORT SERVICE SERVICE MENU...
  • Page 241: Appendix - Maintenance Procedures For The Device Equipment

    NewTom GiANO HR - TECHNICAL MANUAL Appendix - Maintenance procedures for the device equipment In order to ensure the operational safety and functional reliability of the equipment installed, at least once a year. APPLY GREASE (TUTELA MR3) Appendix - Maintenance procedures for the device equipment...
  • Page 242 NewTom GiANO HR - TECHNICAL MANUAL VERIFY WEAR CONDITIONS OF THE BELTS CHECK! Appendix - Maintenance procedures for the device equipment...
  • Page 243 NewTom GiANO HR - TECHNICAL MANUAL Appendix - Maintenance procedures for the device equipment...

Table of Contents