Thermo Scientific MaxQ 2000 CO2 Plus User Manual

Digital co2 resistant shaker
Hide thumbs Also See for MaxQ 2000 CO2 Plus:

Advertisement

Visit us online to register your warranty
www.thermoscientific.com/labwarranty
MaxQ 2000 CO2 Plus
Digital CO2 Resistant Shaker
Catalog #88881101, #88881102
Operation Manual 0000847

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific MaxQ 2000 CO2 Plus

  • Page 1 MaxQ 2000 CO2 Plus Digital CO2 Resistant Shaker Catalog #88881101, #88881102 Operation Manual 0000847 Visit us online to register your warranty www.thermoscientific.com/labwarranty...
  • Page 2 Preface MANUAL NUMBER 0000-847 06/ 06 /2017 Original ECR/ECN DATE DESCRIPTION Thermo Scientific CO2 Resistant Shaker...
  • Page 3 We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products. Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself...over the telephone without a service...
  • Page 4: Table Of Contents

    Warranty Information .........8-1 Thermo Scientific...
  • Page 6: Unpacking And Installation

    2. Check packing list below. If there are any damaged or missing items, contact the manufacturer. 3. Place the instrument on a level and firm surface, to avoid vibration and noise. Table 1-1. Packing List Thermo Scientific 88881101 88881102 88881103 88881104 Cat.
  • Page 7 (88881101), Universal (88881102), Universal Platform and Flask Platform & Flask Clamp Clamp Starter Kit Starter Kit UK Plug Universal Platform (AL) 100ml Flask Clamp 250ml Flask Clamps 500ml Flask Clamps 1L Flask Clamps 2L Flask Clamps CO2 Resistant Shaker Thermo Scientific...
  • Page 8: Universal (Aluminum) Platform Installation

    3. Fasten the universal platform and the fixed plate with four M4x10 cross recessed countersink head screws. Thermo Scientific CO2 Resistant Shaker...
  • Page 9: Rubber Mat Platform Installation

    3. Fasten the platform and the fixed plate with four M4x10 cross recessed countersink head screws. 4. Place the rubber mat on the platform. CO2 Resistant Shaker Thermo Scientific...
  • Page 10: Connections Of The Instrument

    Unpacking and Installation Connections of the Refer to Figure 1-3 when connecting the instrument. Instrument 1) Main body of CO2 resistant shaker 2) Control box 3) Power adapter 4) Power cord Figure 1-3. Circuit Diagram Thermo Scientific CO2 Resistant Shaker...
  • Page 12: Overview

    Noise level ......<60dB Certification ..ROHS, WEEE, CSA, CE Mark Certification ..RoHS, WEEE, cCSAus, CE Mark Power supply ..100-240VAC, 50Hz/60Hz Thermo Scientific CO2 Resistant Shaker...
  • Page 13: Safety Tips

    Caution When storing the cable, pay attention to closing the side plate door, to avoid pinching your fingers. ▲ 1) Main body of CO2 resistant shaker 2) Side plate 3) Control box Figure 2-1. Storing the Signal Cable Diagram CO2 Resistant Shaker Thermo Scientific...
  • Page 14: Maximum Shaking Speed

    ≤1500ml √ 3000ml ≤3000ml √ — ≤2500ml √ — 5000ml ≤5000ml √ — Note The flask clamps (<2L) can be placed anywhere on the universal (aluminum) platform, even if the load is not balanced. ▲ Thermo Scientific CO2 Resistant Shaker...
  • Page 16: Operation

    Key: The power is On/Off. 6) RPM/TIME Key: While instrument is in operation, press to change time and speed display; when setting Key: The instrument is running or stopped. parameter, press to change minutes, hour or speed. Thermo Scientific CO2 Resistant Shaker...
  • Page 17: Speed And Time Settings

    3. In “continuous mode” the LED display will show elapsed run time. 4. After the run is complete, the LED display will read “end”. Press any key to place the shaker in standby mode. CO2 Resistant Shaker Thermo Scientific...
  • Page 18: Alarm System

    WARN indicator light will flash, and the unit will stop running. Power failure recovery When the power supply is cut off suddenly to the shaker, resulting in a power failure, the unit will automatically run at the previous set parameter upon power restoration. Thermo Scientific CO2 Resistant Shaker...
  • Page 20: Application

    CO2 incubator. Refer to Figure 4-1. 1) CO2 Incubator 2) Control Box 3) Power Cord 4) Main body of CO2 Resistant Shaker Figure 4-1. Application in CO2 incubator Thermo Scientific CO2 Resistant Shaker...
  • Page 21 4. The main body of the shaker should be connected to the control box wire through the access port of the CO2 incubator. 5. If the cells are sensitive to vibrations, do not put them in the same incubator or on the shaker. CO2 Resistant Shaker Thermo Scientific...
  • Page 22: Safety Tips And Maintenance

    When the equipment is not functioning properly; • When the equipment is not in use. Maintenance Surfaces can be cleaned with mild detergent and water. Liquid should not be allowed to enter inside the unit. Thermo Scientific CO2 Resistant Shaker...
  • Page 23 2. Remove the platform (with 4 screws). 3. Remove top plate (with 7 screws). 4. Use 70% ETOH to clean the inner area. 5. All allow the unit to dry for one hour, reassemble unit, and operate normally. CO2 Resistant Shaker Thermo Scientific...
  • Page 24: Optional Accessories

    3000ml clamp 88881113 200×200×140 5000ml clamp 88881114 245×245×140 500ml Infusion bottle clamp 88881115 100×100×95 1000ml Infusion bottle clamp 88881116 125×125×125 Funnel Clamp 88881117 Fixed Tube rack 40×Ø14mm 88881118 265x120x140 Fixed Tube rack 40×Ø16mm 88881119 265x120x140 Thermo Scientific CO2 Resistant Shaker...
  • Page 25 Universal (Aluminum) platform 88881122 360x300x5 Rubber Mat platform 88881123 360x305x10 96-well microplate holder 88881124 130x90x15 Spring Wire Rack 88881125 350X240X80 Sticky Mat 88881126 140×140 Control Box 88881127 164×94×27 General Power Adaptor 88881128 US Plug 88881129 CO2 Resistant Shaker Thermo Scientific...
  • Page 26 EU Plug 88881131 UK Plug 88881132 100ml flask clamp x 2 250ml flask clamps x 4 Flask Clamp Starter Kit 88881133 500ml flask clamps x 4 1L flask clamps x 2 2L flask clamp x 2 Thermo Scientific CO2 Resistant Shaker...
  • Page 28: Troubleshooting

    Main-board failure Contact local sales/service The unit has been placed on Place the unit on level and uneven surface sturdy surface. Loud noise Platform loose Fasten platform Accessories loose Fasten accessories Other Keep record for maintenance Thermo Scientific CO2 Resistant Shaker...
  • Page 30: Warranty Information

    Section 8 Warranty Information Thermo Scientific CO2 Resistant Shaker...
  • Page 32 thermoscientific.com © 2014 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.