Summary of Contents for Thermo Scientific MaxQ 2000 CO2 Plus
Page 1
MaxQ 2000 CO2 Plus Digital CO2 Resistant Shaker Catalog #88881101, #88881102 Operation Manual 0000847 Visit us online to register your warranty www.thermoscientific.com/labwarranty...
Page 2
Preface MANUAL NUMBER 0000-847 06/ 06 /2017 Original ECR/ECN DATE DESCRIPTION Thermo Scientific CO2 Resistant Shaker...
Page 3
We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products. Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself...over the telephone without a service...
2. Check packing list below. If there are any damaged or missing items, contact the manufacturer. 3. Place the instrument on a level and firm surface, to avoid vibration and noise. Table 1-1. Packing List Thermo Scientific 88881101 88881102 88881103 88881104 Cat.
3. Fasten the platform and the fixed plate with four M4x10 cross recessed countersink head screws. 4. Place the rubber mat on the platform. CO2 Resistant Shaker Thermo Scientific...
Unpacking and Installation Connections of the Refer to Figure 1-3 when connecting the instrument. Instrument 1) Main body of CO2 resistant shaker 2) Control box 3) Power adapter 4) Power cord Figure 1-3. Circuit Diagram Thermo Scientific CO2 Resistant Shaker...
Noise level ......<60dB Certification ..ROHS, WEEE, CSA, CE Mark Certification ..RoHS, WEEE, cCSAus, CE Mark Power supply ..100-240VAC, 50Hz/60Hz Thermo Scientific CO2 Resistant Shaker...
Caution When storing the cable, pay attention to closing the side plate door, to avoid pinching your fingers. ▲ 1) Main body of CO2 resistant shaker 2) Side plate 3) Control box Figure 2-1. Storing the Signal Cable Diagram CO2 Resistant Shaker Thermo Scientific...
≤1500ml √ 3000ml ≤3000ml √ — ≤2500ml √ — 5000ml ≤5000ml √ — Note The flask clamps (<2L) can be placed anywhere on the universal (aluminum) platform, even if the load is not balanced. ▲ Thermo Scientific CO2 Resistant Shaker...
Key: The power is On/Off. 6) RPM/TIME Key: While instrument is in operation, press to change time and speed display; when setting Key: The instrument is running or stopped. parameter, press to change minutes, hour or speed. Thermo Scientific CO2 Resistant Shaker...
3. In “continuous mode” the LED display will show elapsed run time. 4. After the run is complete, the LED display will read “end”. Press any key to place the shaker in standby mode. CO2 Resistant Shaker Thermo Scientific...
WARN indicator light will flash, and the unit will stop running. Power failure recovery When the power supply is cut off suddenly to the shaker, resulting in a power failure, the unit will automatically run at the previous set parameter upon power restoration. Thermo Scientific CO2 Resistant Shaker...
CO2 incubator. Refer to Figure 4-1. 1) CO2 Incubator 2) Control Box 3) Power Cord 4) Main body of CO2 Resistant Shaker Figure 4-1. Application in CO2 incubator Thermo Scientific CO2 Resistant Shaker...
Page 21
4. The main body of the shaker should be connected to the control box wire through the access port of the CO2 incubator. 5. If the cells are sensitive to vibrations, do not put them in the same incubator or on the shaker. CO2 Resistant Shaker Thermo Scientific...
When the equipment is not functioning properly; • When the equipment is not in use. Maintenance Surfaces can be cleaned with mild detergent and water. Liquid should not be allowed to enter inside the unit. Thermo Scientific CO2 Resistant Shaker...
Page 23
2. Remove the platform (with 4 screws). 3. Remove top plate (with 7 screws). 4. Use 70% ETOH to clean the inner area. 5. All allow the unit to dry for one hour, reassemble unit, and operate normally. CO2 Resistant Shaker Thermo Scientific...
Main-board failure Contact local sales/service The unit has been placed on Place the unit on level and uneven surface sturdy surface. Loud noise Platform loose Fasten platform Accessories loose Fasten accessories Other Keep record for maintenance Thermo Scientific CO2 Resistant Shaker...