MaxQ 2000 CO Plus Digital CO Resistant Shaker Catalog #88881101, #88881102 #88881103, #88881104 Operation Manual 0000847 Visit us online to register your warranty www.thermoscientific.com/labwarranty...
Page 2
Preface MANUAL NUMBER 0000-847 10/28/21 88881127B and 88881181 are introduced. 18/11/2020 Original ECR/ECN DATE DESCRIPTION Thermo Scientific Resistant Shaker...
Page 3
We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products. Whatever Thermo Scientific products you need or use, we will be happy to discuss your applications. If you are experiencing technical problems, working together, we will help you locate the problem and, chances are, correct it yourself...over the telephone without a service...
3. Place the instrument on a level and firm surface, to avoid vibration and noise. Thermo Scientific 88881101 88881102 88881103 88881104 Cat. No. MaxQ 2000 CO2 Plus MaxQ 2000 CO2 Plus Pictutre Promotion Package Promotion Package for MaxQ 2000 CO2 MaxQ 2000 CO2 for USA, includes EMEA & APAC, includes...
Page 7
Platform and Flask Universal Platform & Clamp Starter Kit Flask Clamp Starter Kit EU Plug UK Plug Universal Platform (AL) 100ml Flask Clamp 250ml Flask Clamps 500ml Flask Clamps 1L Flask Clamps 2L Flask Clamps Data Port Resistant Shaker Thermo Scientific...
3. Fasten the platform and the fixed plate with four M4x10 cross recessed countersink head screws. 4. Place the rubber mat on the platform. Thermo Scientific Resistant Shaker...
5) 4-20mA data port Figure 1-4. Circuit Diagram Note: the power first. ▲ 1) When diconnecting the control box from the main body, please cut off 2) No special cable needed to connect with 4-20mA data port. ▲ Resistant Shaker Thermo Scientific...
4. DO NOT clean the unit with flammable or corrosive liquid. 5. Use specified power supply. 6. Maximum load 6kg (13.23 lbs.) @≤300 rpm 7. To ensure proper operation, the unit must be free of surrounding obstructions. Resistant Shaker Thermo Scientific...
1) Main body of CO resistant shaker 2) Slide plate 3) Control box Figure 2-1. Storing the Signal Cable Diagram Orbital Speed vs. Output Current Monitoring System 1) Control box 2) Data port Figure 2-2. Output Current Monitoring System Thermo Scientific Resistant Shaker...
Page 15
Section 2 Overview Orbital Speed vs. Table 2-1. Output current under different speed conditions Output Current Speed (rpm) Output current (mA) Monitoring System (Continued) Resistant Shaker Thermo Scientific...
Table 2-3. Maximum Shaking Speed and Flask Capacity on Universal Platform Flask Size Max Flask Max Speed (rpm) Capacity Size Liquid Volume 50ml ≤50ml 100ml ≤100ml 150ml ≤150ml 200ml ≤200ml 250ml ≤250ml 500ml ≤500ml ≤500ml 1000ml ≤1000ml ≤1000ml 2000ml ≤2000ml ≤1500ml 3000ml ≤3000ml ≤2500ml 5000ml ≤5000ml Thermo Scientific Resistant Shaker...
Page 17
(aluminum) platform, even if the load is not balanced. ▲ 1) The flask clamps (<2L) can be placed anywhere on the 18 inch for the maximum load. ▲ 2) Please refer to the “Liquid Volume” table in the specifications section Resistant Shaker Thermo Scientific...
RPM of the universal (aluminum) platform to the desired RPM setting. Make sure that all the flasks and test tubes are firmly fixed on the universal platform and the flask clamps are fasten securely with the screws. Operations on the Control Panel Thermo Scientific Resistant Shaker...
The minimum unit is 1 minute or 1 hour. 4. During settings, the unit will reset to current values if no value is input after three seconds. 5. Press “Run/Stop” key to run or stop. Resistant Shaker Thermo Scientific...
WARN indicator light will flash, and the unit will stop running. Power failure recovery When the power supply is cut off suddenly to the shaker, resulting in a power failure, the unit will automatically run at the previous set parameter upon power restoration. Thermo Scientific Resistant Shaker...
CO incubator. Refer to Figure 4-1. 1) CO Incubator 2) Control box 3) Power Cord 4) Main body of CO Resistant Shaker Figure 4-1. Application in CO incubator Thermo Scientific Resistant Shaker...
Page 23
4. The main body of the shaker should be connected to the control box wire through the access port of the CO incubator. 5. If the cells are sensitive to vibrations, do not put them in the same incubator or on the shaker Resistant Shaker Thermo Scientific...
• When the equipment is not functioning properly • When the equipment is not in use. Maintenance Surfaces can be cleaned with mild detergent and water. Liquid should not be allowed to enter inside the unit. Thermo Scientific Resistant Shaker...
Page 25
2. Remove the platform (with 4 screws). 3. Remove top plate (with 7 screws). 4. Use 70% ETOH to clean the inner area. 5. All allow the unit to dry for one hour, reassemble unit, and operate normally. Resistant Shaker Thermo Scientific...
Main-board failure Contact local sales/service The unit has been placed on Place the unit on level and uneven surface sturdy surface. Loud noise Platform loose Fasten platform Accessories loose Fasten accessories Other Keep record for maintenance Thermo Scientific Resistant Shaker...