a. CTS Test 2 – Requires DB9F loopback test device
i. CTS setup for scale serial port settings – as on page 4, Steps 1a to 1d.
ii. DB9F Loopback connector - to validate RS232 over ports and adapters.
b. CTS Test 2 – Setup CTS
i. CTS should be setup for as in CTS Test 1 scales test
c. CTS Test 2 – Disconnect Scale from Computer
i. Important – Must first turn off scale and unplug power adapter
ii. Disconnect RS232 cable from scale at computer DB9M COM port.
iii. Place DB9F loopback connector on DB9M on computer
d. CTS Test 2 – COM Validation
i. Turn on CTS COM port – same as done on page 5, steps 2.c,ii
ii. Type a message "Test Loopback" in lower CTS window as below
iii. Press
iv. Test pass - green and yellow message in the upper CTS windows as below
v. Test fail - only the green message - yellow message is missing
e. CTS Test 2 – Test Results Interpretation
i. If CTS Test 2 – COM Validation Pass, COM port link is validated...
ii. Conclusion: CTS Test 2
iii. Conclusion: CTS Test 2
4. Additional information on CTS Test 2 – COM validation
Intelligent Weighing Technology, 4040 Adolfo Road, CA, 93012 USA
E-mail:
sales@intelligentwt.com
1. CTS should be running, if not – run CTS.
a. Be sure all setting are for AJ-MBS as in CTS Test 1
b. Then close CTS COM port - same as done on page 5, steps 2.a.i
button - at right of
1. From application software (CTS) using COM port...
2. ...to Windows O.S. internal COM (and any drivers for VCOM)...
3. ...to physical DB9M COM port on computer
4. ...or to VCOM port used with RS232-to-USB adapter.
1. Hardware, firmware or driver fault in systems or devices below...
a. External RS232-to-USB adapter or Internal VCOM driver
b. External USB port or Internal Windows USB or COM.
1. Likely due to scale settings, or scale serial cable.
2. Recheck scale setting
a. If scale setting correct, fault is likely scale serial port or serial cable
3. Contact your scale service provider for assistance.
Page 14 of 15 - TD0116 RevA01 11-30-2016RP
Web Site:
www.intelligentwt.com
– scale to computer fault is NOT due to...
Pass
– means scale to computer link fault is...
Fail
Phone: +1-805-642-3034
©
Intelligent Weighing Technology 2016.
window
Fax: +1-805-642-4034
Toll Free: 866-920-3000
All rights reserved