HiWeigh HH User Manual

HiWeigh HH User Manual

Physician scale

Advertisement

Quick Links

HH
PHYSICIAN SCALE

Advertisement

Table of Contents
loading

Summary of Contents for HiWeigh HH

  • Page 1 PHYSICIAN SCALE...
  • Page 4 INDEX Introduction Keyboard Functions Technical Description Connecting to load cell Connecting to RS232 Standard Operations On/Off Zero/Tare Units Exchange Height Input and BMI Date/Time Print Further Operations (User Setting) Internal A/D value Hi/Lo/Ok checkweigh function Auto power off Backlight Auto hold function RS232 set (optional) Weighing speed Zero track...
  • Page 5: Keyboard Functions

    1. INTRODUCTION Max. A/D sampling speed: 120 times / second. • Display resolution: from 1/3000 up to 1/30000 • Maximum A/D converting: 24bit • Sensitivity: 0.6mv/V – 3mV/V • Excitation voltage: DC 5V; up to 4 load cells 350 Ω, 8 load cells 700 Ω. •...
  • Page 6: Technical Description

    3. TECHNICAL DESCRIPTION 3.1 Connection of the load cell to indicator INDICATOR LOAD CELL +E (1) ------------- +Excitation -E (5) ------------- -Excitation +S (7) ------------- +Signal -S (8) ------------- -Signal GND (3) ------------- (GND) shield 3.2 Connection of RS232 to PC or Printer INDICATOR PC/PRINTER ---------------- TX...
  • Page 7: Further Operations

    Height Input and BMI Press BMI/Height to enter height input, it display H 170 (default of cm and display A’BC.D” on lb unit), use the keys with arrows to input the height value. Press BMI/Height to confirm the height and it will display the BMI value, such as B 23.6.
  • Page 8: Auto Power Off

    It beeps when the weight is ok (Lo<weight<Hi) It beeps when the weight is out of limitation (Lo>weight or weight>Hi) Auto Power Off UF – 3 Press buttons with arrow to move and change the digits. Press BMI/Height to enter the set of automatic off function– AoFF 00 Auto off deactivated 01-99 Auto off activated in 01-99 minutes, you can change it from 01 minute to 99 minutes...
  • Page 9 Format 1 2 3 4 Communication Protocol UART signal of EIA-RS232 C Data Format 1. Baud Rate : 2400 / 4800 / 9600/19200/38400 BPS 2. Data Bit : 8 BITS 3. Parity Bit : None 4. Stop Bit : 1 BIT Start Bit Data Bit Stop Bit...
  • Page 10 5. Example +1000 pcs stable net weight : s 0D 0A Format 2 (232 4 ~ 6) : Fixed 12 BYTES ASCII (kg g t lb) DATA UNIT CR LF Fixed 15 BYTES ASCII (tl.T lboz) DATA UNIT CR LF Fixed 13 BYTES ASCII (pcs) DATA UNIT...
  • Page 11 Zero Track UF – 8 Press HOLD to change the digits. Press BMI/Height to enter the set of weighing speed - ZP = 1e = 2e = 3e = 4e = 5e * Example: for 1e zero track (if e=20g), the scale will show 0 until it’s more than 20g (1e). Gravity Adjusting UF –...