Operating Standards
6 Operating
Standards
8
TSX Blood Bank Refrigerators
The refrigerators described in this manual are classified for use as stationary
equipment in a Pollution Degree 2 and Overvoltage Category II environment.
These units are designed to operate under the following environmental
conditions:
•
Indoor use
•
Altitude up to 2000 m (6512 feet)
•
Maximum relative humidity 60% for temperatures from 15 to 32°C
(59 to 90°F).
•
For V models: Maximum relative humidity 75% for temperatures
from 10°C to 35°C (50 to 95°F).
•
Main supply voltage fluctuations not to exceed 10% of the nominal
voltage.
•
The refrigerator must not be connected to a GFCI (Ground Fault
Circuit Interrupter) protected outlet as it may be subject to nuisance
tripping.
Operation outside of these conditions will affect the performance of the unit
and the samples stored inside.
Thermo Scientific