7.4.3 Changing Drawer
Thermo Scientific
Drawer slides have a small wire safety clip at the front pilaster which prevent the
Position
slides from falling when the drawer is removed. To change the position of the
drawer slides, complete the following steps:
1. Locate the safety clip.
2. Slip a small screwdriver under the bottom of the wire clip and pry the
clip towards the inside of the refrigerator.
Figure 5. Clip Removal
3. Lift up the slide at the front. The slide is free to move from the front
pilaster.
Figure 6. Front of slide Removal