Installation

7.9 Final Checks

22
TSX Blood Bank Refrigerators
2b. To get an alarm when the switch contacts close, connect the CLOSE
ON FAIL terminal on the cabinet to the CLOSE ON FAIL wire on the
alarm. The COMMON and CLOSE ON FAIL wires must be tied
together in this application.
3. Plug the alarm system service cord into an electrical outlet.
The contacts will trip in the event of a power outage, high temperature alarm or
low temperature alarm.
Before start up, be sure to complete the following steps:
1. Make sure that the unit is free of all wood or cardboard shipping
materials, both inside and outside.
2. Check the positions of the shelves, drawers and baskets. If you want to
adjust the positions, see instructions in Section 7.3, Section 7.4 and
Section 7.5.
3. Verify that the unit is connected to a dedicated circuit.