Table 5-2: Adjustment steps (cont.)
Input
Setup diagram
Ch 2
Sinewave
generator
TDS2000C Series Oscilloscope Service Manual
2-ch model
4-ch model
step number
step number
54
54
55
55
56
56
57
57
58
58
59
59
60
60
61
61
62
62
63
63
64
64
65
65
66
66
67
67
68
68
69
69
70
70
Signal
Signal source setting
Sinewave Generator,
5 kHz, 1 V
50 Ω
Sinewave Generator,
1 MHz, 100 mV
50 Ω
1 MHz, 1 V
Sinewave Generator,
50 Ω
Sinewave Generator,
50 kHz, 1 V
50 Ω
Sinewave Generator,
20 MHz, 1 V
50 Ω
Sinewave Generator,
BWL, 25 mV
50 Ω
Sinewave Generator,
BWL, 50 mV
50 Ω
Sinewave Generator,
BWL, 100 mV
50 Ω
Sinewave Generator,
BWL, 250 mV
50 Ω
Sinewave Generator,
BWL, 500 mV
50 Ω
BWL, 1 V
Sinewave Generator,
50 Ω
Sinewave Generator,
5 kHz, 2.5 V
50 Ω
Sinewave Generator,
1 MHz, 2.5 V
50 Ω
Sinewave Generator,
50 kHz, 2.5 V
50 Ω
Sinewave Generator,
20 MHz, 2.5 V
50 Ω
Sinewave Generator,
BWL, 2.5 V
50 Ω
Sinewave Generator,
BWL, 5 V
50 Ω
Adjustment Procedures
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
5–7
Need help?
Do you have a question about the TDS2022C and is the answer not in the manual?
Questions and answers