AEMC 6416 User Manual page 21

Clamp-on ground resistance tester
Hide thumbs Also See for 6416:
Table of Contents

Advertisement

1.800.561.8187
Menu #
Indication
Setting the Date:
■ Enter the menu by ►. dAtE blinks.
■ Press ▲ or ▼ to select the year. Validate by ►.
■ Press ▲ or ▼ to select the month. Validate by ►.
7
dAtE
■ Press ▲ or ▼ to select the date.
■ Validate and return to the previous menu by ►.
NOTE: In some locations the order of the date values is
displayed as Year, Date, Month.
Setting the Time:
■ Enter the menu by ►. HOUR blinks.
■ Press ▲ or ▼ to select the AM/PM (A. or P.) or 24H
display mode (24H), which blinks. Validate by ►.
8
HOUR
■ Press ▲ or ▼ to select the hour, which blinks.
Validate by ►.
■ Press ▲ or ▼ to select the minutes value, which
blinks.
■ Validate and return to the previous menu by ►.
Selecting Standard or Advanced Operating Mode:
■ Enter the menu by ►. USE blinks.
■ Press ▲ or ▼ to select the Standard or Advanced
mode.
9
USE
- Advanced Mode: The
- Standard Mode: Std is displayed.
■ Validate and return to the previous menu by ►.
Choosing the Test Frequency for the Impedance in
Advanced Mode:
■ Enter the menu by ►. FrEQ blinks.
10
FrEQ
■ Press ▲ or ▼ to select the transposition frequency
of the measured impedance from among the four
possible values: 50, 60, 128, and 2083 Hz.
■ Validate and return to the previous menu by ►.
Clamp-on Ground Resistance Tester Model 6416