8.1 Technical Specifications
All specifications herein mentioned apply to the DSO5000BM series oscilloscopes. Before
checking an oscilloscope from HANTEK to see if it complies with these specifications, make sure
it meets the following conditions:
The oscilloscope must have been operating continuously for twenty minutes under the
specified operating temperature.
The Do Self Cal operation must be performed through the Utility menu if the operating
temperature changes by more than 5 .
The oscilloscope must be within the factory calibration interval.
All specifications are guaranteed unless noted 'typical'.
Oscilloscope Specifications
Horizontal
Sample Rate Range
Waveform Interpolation
Record Length
SEC/DIV Range
Sample Rate and
Delay Time Accuracy
Delta Time Measurement
Accuracy
(Full Bandwidth)
Position Range
DSO5000 Series Digital Storage Oscilloscope User Manual
Chapter 8 Specifications
℃
1GS/s
(sin x)/x
Maximum 2M samples per single-channel; maximum 1M
samples per dual-channel (4K,16K,40K optional)
DSO5062BM
DSO5102BM
4ns/div to 40s/div, in a 2, 4, 8
sequence
±50ppm over any ≥1ms time interval
Single-shot, Normal mode
± (1 sample interval +100ppm × reading + 0.6ns)
>16 averages
± (1 sample interval + 100ppm × reading + 0.4ns)
Sample interval = s/div ÷ 200
DSO5062BM DSO5102BM
200μs/div to 40s/div
DSO5202BM
2ns/div to 10ns/div
Specifications
DSO5202BM
2ns/div to 40s/div, in a 2, 4, 8
sequence
(-8div × s/div) to 400s
(-4div × s/div) to 20ms