3.6 Self Calibration

The self calibration routine helps optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but should always run it if the
ambient temperature changes by 5
the oscilloscope and wait for 20 minutes until it has adequately warmed up.
To compensate the signal path, disconnect any probes or cables from the front-panel input
connectors. Then, push the UTILITY button, select the Do Self Cal option and follow the
directions on the screen.
.
DSO5000 Series Digital Storage Oscilloscope User Manual
or more. For a more accurate calibration, please power on
Getting Started Guide
10