Specifications - DeVilbiss IntelliPAP DV54 Series Manual

Positive airway pressure device
Hide thumbs Also See for IntelliPAP DV54 Series:
Table of Contents

Advertisement

Available languages

Available languages

issue
Air from flow
generator seems to
be too warm.
Nasal, sinus or ear
pain, runny nose
Heater setting is
not visible on
display

speciFicatioNs

cpap
Size ........................................................................................................... 4.2"H x 6.5"W x 6.9"D (10.7cm x 16.5cm x 17.5cm)
Weight ............................................................................................................................................................. 1.22 kg (2.7 lbs.)
Electrical Requirements AC ...................................................................................................................... 100-240V~, 50/60 Hz
Electrical Requirements DC .................................................................................................................. 10.4 - 15 VDC, 5 Amps
Maximum Power Consumption ......................................................65 watts max from AC power source (flow generator only)
Pressure Range ......................................................................................................................................................3-20 cmH
Operating Temperature Range ....................................................................................................... 41˚F to 104˚F (5˚C to 40˚C)
Operating Humidity Range .......................................................................................................0% to 95% RH non-condensing
Operating Atmospheric Conditions ..........................................................................................Sea level to 9,000 feet (2743 m)
Storage & Transportation Temperature Range ..................................................................... -40°F to +158°F (-40°C to 70°C)
Storage & Transportation Humidity Range ...............................................................................0% to 95% RH non-condensing
Maximum Limited Pressure ...........................................................................................................20 cmH
Sound Pressure Level (tested per ISO 17510) ...............................................................................................................26 dBA
Sound Power Level .........................................................................................................................................................38 dBA
Humidifier
Humidity Output (in the operating flow range): ................................................................................................ ≥10 mgH
Size: ......................................................................................................2.6" H x 6.3" W x 8.4" D (6.6 cm x 16.0 cm x 21.3 cm)
Weight: ........................................................................................................................................................1.75 lbs. (0.794 Kg)
Power Consumption: .................................................................................................................................................... 85 Watts
DV5HH Electrical Requirements ................................................................................................................100-240VAC, 0.95 A
Heater Plate Thermal Cutout:..............................................................................................................................221°F (105°C)
Power/Temperature Control: ......................................................................................Setting of 1 (minimum) to 10 (maximum)
Heater Plate Temperature: ...............................................................................84°F to 149°F (approx. 29°C to approx. 65°C)
Operating Atmospheric Pressure Range: ........................................................................................................70.0 – 106.0 kPA
Filter specifications
Standard Filter .........................................................................................................................................> 3.0 micron particles
Optional Fine Particle Filter .....................................................................................................................> 0.3 micron particles
a-DV54
possible cause
1. Air filters are dirty.
2. Air inlet port is blocked.
3. Room temperature is too high.
4. Device is located near a heat source.
5. Humidifier (optional) heater setting is too
high.
You may have a reaction to air flow pressure. Discontinue use and contact physician.
1. The flow generator is not fully seated onto
the humidifier cradle.
2. Your CPAP device is not equipped with the
optional humidifier.
remedy
1. Clean filters.
2. Unblock air inlet.
3. Lower room temperature.
4. Locate device away from heat source.
5. Lower heater setting.
1. Ensure proper contact is made (Fig. 3).
O
2
O under normal use
2
O/l air
2
eN - 13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Intellipap dv54Intellipap dv5hh

Table of Contents