Connect the quick-disconnect to the dental office water
supply or a Cavitron DualSelect Dispensing System.
Inspect all connections to make certain there are no
leaks.
To remove the water line from the Cavitron Plus System,
turn off the dental office water supply. Disconnect the
water supply line from the dental office water supply. If
a quick-disconnect connector is attached to the end of
the hose, relieve the water pressure by pressing the tip
of the connector in an appropriate container and allow
water to drain. To remove the hose from the system,
push on the outer ring of the system's water inlet and
gently pull out the water line.
7 7 Tap-On
Foot Pedal Battery
TM
Installation/Replacement
Turn Tap-On
Wireless Foot Pedal over and using a
TM
Philips screwdriver carefully remove battery cover screw
and battery cover. If applicable, remove used batteries
and install two new "AA" batteries as shown. Do not
depress Tap-On
Foot Pedal while installing batteries.
TM
The communication light will blink for approximately
two seconds to indicate the Tap-On
to communicate with the unit. If the light does not blink,
check the batteries. If the batteries are good and the
light doesn't blink, a communications error may exist.
To re-establish communication with Tap-On
review Synchronization procedure, section 7.10.
The remote frequency communication can be bypassed
using the auxillary Tap-On
Section 11.2 Technical Support and Repair for further
action.
Replace the battery cover and screw and hand tighten
cover with Philips screwdriver.
Remove batteries if Tap-On
for an extended period of time.
7 8 Tap-On
Foot Pedal Synchronization
TM
The Tap-On
Wireless Foot Pedal supplied with your
TM
system has been factory synchronized with the base unit.
Should a replacement Tap-On
synchronization will be required prior to system operation.
Perform the following steps to synchronize the Tap-On
Pedal with the base unit.
1.
Turn the MainPower switch located at the center
Look for blinking
communications
light.
Foot Pedal's ability
TM
Foot Pedal
TM
Foot Pedal cable. Refer to
TM
Foot Pedal is to be stored
TM
Foot Pedal be necessary,
TM
TM
front underside of the system to the OFF (O) position.
2.
Install a new set of "AA" batteries into the foot
control (see section 7.9). Leave the battery cover of
the Tap-On
is accessible.
3.
Maintain a distance of no more than 10 feet (3
meters) between the base unit and Tap-On
Pedal during the synchronization process.
4.
Remove any inserts from the handpiece and adjust
the Power Level Control out of Rinse Mode. Turn the
Main Power switch to the ON (I) position and wait for
the Diagnostic Display graphics to light (refer to
Section 8.2).
5.
While all graphics are lit, press the Purge button,
located on the Diagnostic Display.
PURGE
The graphics will begin to blink in a sequential
pattern, representing the synchronization mode.
This mode will last 5 to 6 seconds.
6.
During this mode, press the red button located in the
battery compartment of the Tap-On
will complete the synchronization process.
7.
Synchronization is successful when all graphic lights
blink at the same time.
8.
To verify proper communication, press the foot
control to the Boost position (Tap-On
fully depressed – 2
graphic on base unit illuminates.
9.
Attach battery cover and tighten the screw.
10.
In the event communication cannot be established,
temporarily use the supplied Auxillary Tap-On
Pedal Cable to connect the Tap-On
directly to the unit.
Foot
10
Foot Pedal open so the red push button
TM
TM
TM
position) and ensure the Boost
nd
Foot Pedal
TM
Foot
TM
Foot Pedal. This
Foot Pedal
Foot
TM

Advertisement