Fuse Replacement Instructions

Replacing the fuses:
Replacement Fuses:
115V: 0.10A, 250V rated Slo-Blo
230V: 0.05A, 250V rated Slo-Blo
NOTE: The Calamus
fuses installed for a 115VAC power source. If power source is 230
VAC, ensure that 250V 0.05A rated fuses are installed.
Warning: Turn the power off and unplug the unit before following
the fuse replacement steps.
1. Remove the fuseholder from the power inlet connector. See Fig. 6.
2. Replace the fuses in the fuseholder.
3. Replace the fuseholder.
Fuse Holder
®
type (Fuse size: 5 x 20mm)
®
type (Fuse size: 5 x 20mm)
®
Pack is manufactured with 250V 0.10A rated
Figure 6
Fuse
Replacement
AC Power
Inlet Connector
13