Storz 11301D4 Instructions For Use Manual page 17

Led battery light source with fast screw thread
Table of Contents

Advertisement

Test fre-
quency
MHz
1,720
1,845
1,970
2,450
5,240
5,500
5,785
8.4 Table 3 – Test levels for radiated and conducted
immunity tests
Guidelines and manufacturer's declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The user
of the product should make sure that it is used in such an environment.
Interference immunity
tests
Conducted HF distur-
bances acc. to IEC
61000-4-6
Radiated HF distur-
bances acc. to IEC
61000-4-3
Instructions for use • LED battery light source with fast screw thread • KCC914_EN_V1.0_11-2020_IFU_CE-
MDR
Frequency
Radio service Modulation
band
MHz
1,700 – 1,990 GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE band 1, 3,
4, 25,
UMTS
2,400 – 2,570 Bluetooth,
WLAN 802.11
b/g/n,
RFID 2450,
LTE band 7
5,100 – 5,800 WLAN 802.11
a/n
EN/IEC 60601 test
level
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Electromagnetic compatibility
Immunity
test level
V/m
Pulse modula-
28
tion
217 Hz
Pulse modula-
28
tion
217 Hz
Pulse modula-
9
tion
217 Hz
Compliance
Electromagnetic envi-
level
ronment – guidelines
3 V
Portable and mobile HF
rms
communications equip-
ment should be used no
closer to any part of the
3 V/m
product, including cables,
than the recommended
separation distance calcu-
lated from the equation
applicable to the fre-
quency of the transmitter.
Recommended separation
distances:
d = 1.2 √P
Where P is the rated
power of the transmitter in
watts [W] according to the
information provided by
the transmitter manufac-
turer and d is the recom-
mended separation dis-
tance in meters [m].
Compliance
level
V/m
28
28
9
17