Download Print this page

Specifications - Sanwa CAM600S Instruction Manual

Analog clamp meter

Advertisement

【5】SPECIFICATIONS
5-1 Measurement Range and Accuracy
(23℃±5℃, 80%RH max. No condensation)
Function
ACA
ACV
DCV
Ω
(Temp.)
5-2 Others
・Max. clamp size :φ36mm or 10×50mm
・Meter
・AC rectification
・Circuit protection : The circuit is proected by fuse even when voltage of up to AC 230V
*Built-in battery
・Internal fuse
・Applicable circuit voltage : AC 600V or less
・Applicable standard : IEC 1010-2 Installation Category (Overvoltage Category)Ⅲ.
・Withstand voltage : AC 5550V between iron core and rear case (1min.)
・Applicable altitude : Up to 2000m(above sea level)
・Operating temperature/humidity range
・Storage temperature/humidity range
・Dimensions and weight : 221 (H) ×97 (W) ×43 (D) mm・420g
・Accessories
・Optional accessories : Temperature probe (model THP)
Range
6−15−60−150−600
150−300−600
60
1k − 100k
(×1) (×100)
(−10〜200℃)
: Moving coil type. 183μA
: Half-wave rectification
is impressed on each range for 5 seconds.
: R6 (IEC) or UM-3 1.5V×1
: 0.5A/250V, φ6.3×30mm Fast acting fuse.
Pollution Degree Ⅱ
: 0〜40℃, 80% RH max. no condensation
: −10〜+50℃, 70% RH max. no condensation
: Test lead (TL21) 1. Carrying case (C-CAM6) 1. Instruction manual 1.
Spare fuse (0.5A/250V, φ6.3×30mm) 1.
Tolerance
±3%
against f.s.
300A or more
±4% against f.s.
±3%
against f.s.
±3%
against f.s.
±3%
scale length
±3.5%
scale length
Remarks
sine wave
50,60Hz
center 30Ω-3kΩ
battery 1.5V×1
with optional probe
(model THP)

Advertisement

loading