Specifications; Sanwa Web Site; General Specifications - Sanwa LCR700 Instruction Manual

Digital lcr meter
Table of Contents

Advertisement

information. An instrument sent to SANWA / agent / distributor
without above information will be returned to the customer.
Note:
1) Prior to requesting repair, please check the following:
Capacity of the built-in battery, polarity of installation and
discontinuity of the test leads.
2) Repair during the warranty period:
The failed meter will be repaired in accordance with the
conditions stipulated in 8-1 Warranty and Provision.
3) Repair after the warranty period has expired:
In some cases, repair and transportation cost may become
higher than the price of the product. Please contact SANWA
authorized agent / service provider in advance.
The minimum retention period of service functional parts is 6
years after the discontinuation of manufacture. This retention
period is the repair warranty period. Please note, however,
if such functional parts become unavailable for reasons of
discontinuation of manufacture, etc., the retention period may
become shorter accordingly.
4) Precautions when sending the product to be repaired:
To ensure the safety of the product during transportation, place
the product in a box that is larger than the product 5 times or
more in volume and fill cushion materials fully and then clearly
mark "Repair Product Enclosed" on the box surface. The cost
of sending and returning the product shall be borne by the
customer.

7-3 SANWA web site

http://www.sanwa-meter.co.jp
E-mail: exp_sales@sanwa-meter.co.jp
― 35 ―

[8] SPECIFICATIONS

8-1 General Specifications
Main display 20,000 counts: Ls / Lp / Cs / Cp / Rs / Rp / DCR
LCD display
(with Backlight)
Sub display 2,000 counts: D/Q/θ/ESR/RP
Over-range
While the meter is reading outside limits, numeric
part of the display indicates [OL].
indication
Battery power
4 levels
indication
Measurable
Ls / Lp / Cs / Cp / Rs / Rp / DCR
(including D/Q/θ/ESR/RP)
parameters
Measuring
Series / Parallel
mode
Range
Automatic
selection
Measuring
4 terminals plus guard shielding
terminals
Accepts banana plugs.
Auto LCR
L: 20.000 μH ~ 20.000 kH
C: 200.00 pF ~ 20.00 mF
measurement
ranges
R: 20.000 Ω ~ 200.0 MΩ
DCR
measurement
R: 200.00 Ω ~ 200.0 MΩ
ranges
Measuring
100 Hz / 120 Hz / 1 kHz / 10 kHz / 100 kHz
frequencies
Measurement
1.2 times / second (LCR mode)
rate
0.5 times / second (DCR mode)
Selectable
±0.25 %, ±0.5 %, ±1 %, ±2 %, ±5 %, ±10 %, ±20 %,
-20 % ~ +80 %
tolerances
Measuring
0.63 Vrms (nominal), 0.9 Vdc (nominal)
signal level
Operating
Altitude: < 2,000 m Pollution degree: II
conditions
Specification
guaranteed
18 ~ 28 ℃ < 80 % RH
temperature/
humidity
― 36 ―