Refer to the following table if the station fails to operate properly. If the
suggested corrective actions do not solve the problem, contact Labconco
for additional assistance.
PROBLEM
Station blower
won't turn on
Filter System Light
turns on
18
:
:
o
o
t
i
n
g
h
o
o
t
i
n
g
CAUSE
Unit not plugged
into outlet
Circuit breaker
tripped
Blockage of the
prefilter
HEPA filter is
blocked
CORRECTIVE ACTION
Plug the station into appropriate
electrical service.
Check connection to the top of
station.
Reset circuit breaker.
Check the prefilter; replace if
needed. Ensure that there is no
obstruction of the exhaust on top of
the station.
If replacing prefilter does not help,
replace the HEPA filter.