Description Of 5004A Signature Analyzer; Signature Analysis - HP 5004A Operating And Service Manual

Signature analyzer
Table of Contents

Advertisement

Model 5004A
General Information
DISPLAY:
Signature: Four-digit hexadecimal.
GATE, UNSTABLE SIGNATURE indicators: Panel Lights. Stretching: 100 millisconds.
Probe-tip indicator: Light indicates high, low, bad-level, and pulsing states.
PROBABILITY OF CLASSIFYING CORRECT DATA STREAM A S CORRECT: 100%.
PROBABILITY OF CLASSIFYING FAULTY DATA STREAM A S FAULTY: 99.998%.
MINIMUM GATE LENGTH: One clock cycle.
MINIMUM TIMING BETWEEN GATES (from last STOP to next START): One clock cycle.
DATA PROBE:
lnput Impedance: 50
Threshold: Logic one: 2.0 Volt +.2 -.3.
Setup Time: 15 nanoseconds, with .2 volt over-drive. (Data required to be valid at least
15 nanoseconds before selected clock edge.)
GATING INPUT LINES:
START, STOP, CLOCK inputs: lnput Impedance: 50 K f l to 1.4 volt, nominal. Shunted by
7 pF, nominal.
START, STOP inputs:
Setup Time: 25 nanoseconds. (START, STOP to be valid at least 25 nanoseconds before
selected clock edge.)
CLOCK INPUT:
Minimum Clock Time in High or Low State: 50 nanoseconds.
VOLTAGE OVERLOAD PROTECTION: All inputs +I50 volts continuous.
OPERATING ENVIRONMENT:
Temperature: 0-55OC.
Relative Humidity: 95% at 40°C.
Altitude: 4,600M.
POWER REQUIREMENTS:
Option 100: l00V ac line, +5%, -lo%, 48-440 Hz
Option 120: 120V ac line, +5%, -lo%, 48-440 Hz
Option 220: 220V ac line, +5%, -lo%, 48- 66 Hz
Option 240: 240V ac line, +5%, -lo%, 48- 66 Hz
WEIGHT: Net: 2.5 kg, 5.5 lbs.
DIMENSIONS:
90 mm high x 215 mm wide x 300 mm deep (3l/, in. x 5l/, in. x 12 in.)
Dimensions exclude tilt bale, probes, and pouch.
1-15.
DESCRIPTION O F 5004A SIGNATURE ANALYZER
1-16. The HP Model 5004A Signature Analyzer is a test instrument for troubleshooting com-
plex electronic logic circuits. It uses the signature analysis technique of troubleshooting.
1-17.

Signature Analysis

1-18. Signature analysis is a method of troubleshooting complex electronic logic circuits to
the individual component level. To use signature analysis with the 5004A, the unit to be tested
must have certain characteristics included with the original design. Typically a logic product
Table 7-7. Specifications
Characters 0,1,2,3,4,5,6,7,8,9,A,C,F,H,P,U.
Minimum pulsewidth: 10nanoseconds.
to 1.4 Volt, nominal. Shunted by 7 pF, nominal.
Logic zero: .8 Volt, +.3 -.2.
Threshold: 1.4 volt k.6 (.l
Shipping: 7.7 kg, 17 lbs.
Stretching: 50milliseconds.
volt hysteresis, typical).
k 250 volts intermittent.
250 volts ac for 1 minute.

Advertisement

Table of Contents
loading

Table of Contents