Routine Maintenance; Cleaning The Microtymp 4; Eartip And Probe - Welch Allyn MICROTYMP 4 User Manual

Hand held portable tympanometer
Table of Contents

Advertisement

Welch Allyn MicroTymp® 4 Tympanometer User Manual
R
M
OUTINE
AINTENANCE
C
M
LEANING THE
ICRO
WARNING Electric shock hazard. Before cleaning the device, disconnect the power cord
from the power source and the device.
WARNING Take care to prevent water or other fluid from entering any connectors on
the device. Should this occur, dry the connectors and check the accuracy of all operating
functions.
CAUTION The device is not heat-resistant. Do not autoclave.
The MicroTymp 4 is a precision instrument. Handle it carefully to ensure its continued accuracy
and service. Use a soft damp cloth and mild detergent to clean the instrument panel and case
when required. Ensure no moisture enters the instrument.
If low-level disinfection is required, Oxivir Tb (Diversey, Inc.) a hydrogen-peroxide based solution,
was tested and found to be compatible with the plastic device housing.
E
P
ARTIP AND
ROBE
WARNING
Handle the probe and accessories with care. Do not allow moisture, condensation,
fluids or debris to enter the probe.
DIR 80024157 Ver A
8000
9000
10,000
T
4
YMP
2438.4
2743.2
3048.0
2.7 ±0.1
2.8 ±0.1
2.9 ±0.1
Page 30