Final Tests - Siemens MAMMOMAT Novation Maintenance Instructions Manual

Table of Contents

Advertisement

Miscellaneous
6.4

Final tests

Remove the measurement devices.
PMF
Operating problems
Were there any operating problems during the checks?
PMP
Covers
Remove protective strips and attach all removed covers. Check that all covers are com-
plete and firmly seated, including the cable duct.
SIE
Protective conductor test
Perform the protective conductor test according to ARTD-002.731.17....
PMP
Cleaning / Damaged paint
Clean the unit using the materials recommended in the Instructions for use and repair
any damaged paint.
System ON.
QSQ
Final test exposure
Take one final test exposure.
Siemens AG
Medical Solutions
SPB7-250.831.01.02.02
08.04
MAMMOMAT Novation
Page 31 of 32
CS SD 24
31