Miscellaneous
6.4
Final tests
•
Remove the measurement devices.
PMF
Operating problems
•
Were there any operating problems during the checks?
PMP
Covers
•
Remove protective strips and attach all removed covers. Check that all covers are com-
plete and firmly seated, including the cable duct.
SIE
Protective conductor test
•
Perform the protective conductor test according to ARTD-002.731.17....
PMP
Cleaning / Damaged paint
•
Clean the unit using the materials recommended in the Instructions for use and repair
any damaged paint.
•
System ON.
QSQ
Final test exposure
•
Take one final test exposure.
Siemens AG
Medical Solutions
SPB7-250.831.01.02.02
08.04
MAMMOMAT Novation
Page 31 of 32
CS SD 24
31