Evaluating Test Values - Siemens MAMMOMAT Novation Maintenance Instructions Manual

Table of Contents

Advertisement

24
5.1

Evaluating test values

PMF
Anode acceleration
Listen to the sound generated by the accelerating anode. Is the sound normal?
PMF
kV and mA
Do the measured kV and mA values agree with the values selected?
PMF
mAs selection
Do the measured mAs values agree with the values selected?
PMF
mAs value displayed
Does the mAs value measured agree with the mAs value displayed (for AEC)?
PMF
Grid voltage
Attach the magnification table. Measure the grid voltage (small focus) on tube test point
G against test point H
col?
Test both focal spots (if available).
PMF
Signal lamp
Does the radiation-ON indicator light up?
PMF
Chest wall missed tissue
Perform the check according to the Quality Control Manual, chapter "Chest wall missed
tissue", to ensure that the chest wall missed tissue is within the tolerance of max. 5 mm.
X
PMF
Spatial resolution
Perform the check according to the Quality Control Manual, chapter "Spatial resolu-
tion", to ensure that a sufficient spatial resolution is obtained with the system.
X
PMF
Mean glandular dose
Perform the check according to the Quality Control Manual, chapter "Mean glandular
dose, to ensure that the mean glandular doses are according to regulations.
X
MAMMOMAT Novation
. Does the value measured agree with the value in the test proto-
3
SPB7-250.831.01.02.02
08.04
Test exposures
Page 24 of 32
CS SD 24
Siemens AG
Medical Solutions